The eCareManager system. Software intended for use in data collection, storage and clinical information management with independent bedside devices, and ancillary systems that are connected either directly or through networks. For use in a hospital environment. eCareManager allows clinicians to generate a PDF display of an order (new, change, discontinue), which is printed at the bedside and/or hospital pharmacy. This printed PDF should be reviewed by the clinician and verified by the hospital pharmacist and entered into the hospital pharmacy system.
Enforcement
- Recall Number
- Z-0399-2016
- Event ID
- 70931
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Visicu, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 16, 2015
- Initiation Date
- March 27, 2015
- Classification Date
- December 10, 2015
- Termination Date
- September 23, 2016
- Address
- 217 E Redwood St Ste 1900, N/A, Baltimore, MD, 21202-3315, United States
Description
The eCareManager system. Software intended for use in data collection, storage and clinical information management with independent bedside devices, and ancillary systems that are connected either directly or through networks. For use in a hospital environment. eCareManager allows clinicians to generate a PDF display of an order (new, change, discontinue), which is printed at the bedside and/or hospital pharmacy. This printed PDF should be reviewed by the clinician and verified by the hospital pharmacist and entered into the hospital pharmacy system.
A software defect may cause incorrect medication order change. If the user decides to edit the order prior to electronic signature AND selects the return to Previous Screen function, the previously selected medication is replaced with whatever medication is in the number one spot on the complete medication list. At the stage where the order is signed, the medication displayed may be incorrect.
eCareManager v3.7, 3.7.1, v3.8, and v3.9
US distribution only.