FDA Enforcement Class II Terminated

TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic endoscopic procedures, to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic / laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions. It is comprised of a single patient use, bladeless Obturator designed to allow penetration and positioning at the required site within the abdomen, and several single use Cannulas, which serve as working channels for the procedure. The Cannulas are pre-loaded with two synthetic absorbable sutures, each attached to a synthetic absorbable anchor. The deployment of the anchors and sutures is achieved by utilizing a mechanism within the Obturator. The Cannula is 115 mm long and is compatible for use with laparoscopic devices of 5 - 12 mm outer diameter.

Recall: Z-0398-2021 · Reported November 11, 2020

Enforcement

Recall Number
Z-0398-2021
Event ID
86483
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Gordian Surgical
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 11, 2020
Initiation Date
September 22, 2020
Classification Date
November 3, 2020
Termination Date
February 11, 2021
Address
Tchelet 17, Trendlines Building, Misgav, N/A, Misgav, N/A, N/A, Israel

Description

TroClose 1200 - Product Usage: intended for use in a variety of gynecologic, general and urologic endoscopic procedures, to create and maintain a port of entry and to facilitate the delivery of absorbable sutures and anchors through soft tissues of the body during endoscopic / laparoscopic surgery. The trocar may be used with or without visualization for primary and secondary insertions. It is comprised of a single patient use, bladeless Obturator designed to allow penetration and positioning at the required site within the abdomen, and several single use Cannulas, which serve as working channels for the procedure. The Cannulas are pre-loaded with two synthetic absorbable sutures, each attached to a synthetic absorbable anchor. The deployment of the anchors and sutures is achieved by utilizing a mechanism within the Obturator. The Cannula is 115 mm long and is compatible for use with laparoscopic devices of 5 - 12 mm outer diameter.

Reason

Endotoxin test results found above the acceptable levels in LAL test (above 20EU/device).

Code Info

Catalog Number: GSTC1200 Lot Numbers: AS200009/3, Exp 14 FEB 2022 AS200012/4, Exp 23 MAR 2022

Distribution

US Nationwide distribution including in the states of Texas and Michigan.

Quantity

300 total devices