FDA Enforcement Class I Terminated

Laparoscopy Gyn Pack, Catalog number 900-605. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Recall: Z-0390-2015 · Reported December 10, 2014

Enforcement

Recall Number
Z-0390-2015
Event ID
69448
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Customed, Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 10, 2014
Initiation Date
October 8, 2014
Classification Date
November 28, 2014
Termination Date
August 14, 2017
Address
Calle Igualdad Final #7, N/A, Fajardo, PR, 00738, United States

Description

Laparoscopy Gyn Pack, Catalog number 900-605. Used by medical staff to perform patient care procedures such as wound healing, suture removals, irrigate wounds, etc.

Reason

Significant compromised sterility, not limited to potentially damaged packaging. Serious deficiencies in the manufacturing processes and uncontrolled and inadequate storage conditions increase the risk of contamination and may result in patient infection.

Code Info

Lot numbers 140513375 140613772 140714301 140814679 140915208

Distribution

US Distribution to Florida, New York and Puerto Rico..

Quantity

160