FDA Enforcement
Class II
Terminated
DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.
Recall: Z-0388-2019
·
Reported November 28, 2018
Enforcement
- Recall Number
- Z-0388-2019
- Event ID
- 81454
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- GE Healthcare Biosciences
- Voluntary / Mandated
- FDA Mandated
- Initial Notification
- Letter
- Report Date
- November 28, 2018
- Initiation Date
- September 10, 2018
- Classification Date
- November 20, 2018
- Termination Date
- May 12, 2020
- Address
- 14 Walkup Dr, N/A, Westborough, MA, 01581-1019, United States
Description
DeltaVision OMX SR Imaging System Model Number: 29115476 The DeltaVision OMX SR Imaging System is a piece of laboratory equipment for super resolution imaging offluorescently-la beling biological samples. Techniques such as 3D Structured Illumination, Total Internal Reflection Fluorescence and localization microscopy may be used to examine a variety of samples over long periods of time.
Reason
Non-conformance with the DeltaVision OMX SR, the interlocks are not wired in a redundant fashion to make the interlock system single-fault tolerant.
Code Info
System Serial number OM20072 - OM40122
Distribution
Worldwide - US Nationwide Distribution
Quantity
39