FDA Enforcement Class III Terminated

Single Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only. store at 2 - 8 degrees C. Contents: Contains 13 mg/Ml of Norbuprenorphine in human urine with sodium azide as a preservative. Manufacturer: Lin-Zhi International, Inc. 670 Almanor Ave. Sunnyvale, CA 94085 The Norbuprenorphine Drug of Abuse (DUB) Positive control is used as assayed quality control material to monitor the precision of the Lin-Zhi International, Inc. Buprenorphine Enzyme immunoassay on the Synchron clinical analyzer. It in intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine.

Recall: Z-0386-2014 · Reported December 4, 2013

Enforcement

Recall Number
Z-0386-2014
Event ID
66542
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Lin-Zhi International Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 4, 2013
Initiation Date
October 3, 2013
Classification Date
November 25, 2013
Termination Date
December 26, 2013
Address
670 Almanor Ave, N/A, Sunnyvale, CA, 94085-3513, United States

Description

Single Analyte Urine DAU Control (5 m L Norbuprenorphine, Level II control For In Vitro Diagnostic Use Only. store at 2 - 8 degrees C. Contents: Contains 13 mg/Ml of Norbuprenorphine in human urine with sodium azide as a preservative. Manufacturer: Lin-Zhi International, Inc. 670 Almanor Ave. Sunnyvale, CA 94085 The Norbuprenorphine Drug of Abuse (DUB) Positive control is used as assayed quality control material to monitor the precision of the Lin-Zhi International, Inc. Buprenorphine Enzyme immunoassay on the Synchron clinical analyzer. It in intended for the qualitative and semi-quantitative determination of norbuprenorphine in human urine.

Reason

Product catalog # A68825 Norbuprenorphine 13 ng/mL Level 2 Controls gives a decreased concentration reading due to the degradation of the NBUP analyte in solution.

Code Info

Catalog #A68825, Lot #1307086; Exp. 3/28/2014/

Distribution

US Distribution in CA.

Quantity

60 items