LMA MADett, Endotracheal tube Mucosal Atomization Device, Rx only, Product Usage: These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex Medical is recalling these products as they may produce a straight stream instead of a fully atomized plume of medication. It is unlikely that serious adverse health consequences will occur in the event of a failure to deliver an atomized plume; however, this may result in inadequate topical anesthesia which may lead to some discomfort, further attempts to deliver topical anesthesia, or the use of alternative methods of anesthesia.
Enforcement
- Recall Number
- Z-0385-2017
- Event ID
- 75554
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Teleflex Medical
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 28, 2016
- Initiation Date
- October 27, 2016
- Classification Date
- December 22, 2016
- Termination Date
- September 25, 2020
- Address
- 2917 Weck Dr, N/A, Research Triangle Park, NC, 27709-0186, United States
Description
LMA MADett, Endotracheal tube Mucosal Atomization Device, Rx only, Product Usage: These products are used for the delivery of topical anesthesia via an atomized spray to the oral, nasal, pharyngeal or laryngeal mucosa. Teleflex Medical is recalling these products as they may produce a straight stream instead of a fully atomized plume of medication. It is unlikely that serious adverse health consequences will occur in the event of a failure to deliver an atomized plume; however, this may result in inadequate topical anesthesia which may lead to some discomfort, further attempts to deliver topical anesthesia, or the use of alternative methods of anesthesia.
Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.
Product Code: MAD710, Lot# 160120.
Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
2,674,650 units in total