ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
Enforcement
- Recall Number
- Z-0385-2014
- Event ID
- 66605
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Roche Diagnostics Operations, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 4, 2013
- Initiation Date
- October 25, 2013
- Classification Date
- November 22, 2013
- Termination Date
- January 14, 2015
- Address
- 9115 Hague Rd, N/A, Indianapolis, IN, 46256-1025, United States
Description
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.
AccuChek Aviva Combo meter 05075645005 packaged in Accu-Chek Combo Kit mg US/English version, 05458544001 and Accu-Chek Combo Kit mg US/Spanish version, 05504686001.
Nationwide Distribution.
6,595 units