FDA Enforcement Class II Terminated

Bartels Legionella Urinary Antigen- an adjunct to culture for the presumptive diagnosis of past or current Legionnaire's Disease by qualitative detection of Legionella Pneumophila seragroup 1 antigen in human urine. Product Code: B1029-440

Recall: Z-0382-2022 · Reported December 22, 2021

Enforcement

Recall Number
Z-0382-2022
Event ID
89088
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Clark Laboratories, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 22, 2021
Initiation Date
October 11, 2021
Classification Date
December 15, 2021
Termination Date
March 9, 2023
Address
2823 Girts Rd, Jamestown, NY, 14701-9666, United States

Description

Bartels Legionella Urinary Antigen- an adjunct to culture for the presumptive diagnosis of past or current Legionnaire's Disease by qualitative detection of Legionella Pneumophila seragroup 1 antigen in human urine. Product Code: B1029-440

Reason

Contamination of the conjugate may cause a pipetting issue which then could lead to an invalid assay

Code Info

Lot # 056 Expiration Date: 2022-11-28 UDI: 05391516744065

Distribution

OH, UT

Quantity

53 units