FDA Enforcement Class II Terminated

SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Recall: Z-0382-2013 · Reported November 28, 2012

Enforcement

Recall Number
Z-0382-2013
Event ID
63386
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Spine
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 28, 2012
Initiation Date
August 9, 2012
Classification Date
November 19, 2012
Termination Date
September 30, 2015
Address
2 Pearl Ct, N/A, Allendale, NJ, 07401-1611, United States

Description

SPECIALTY AUDIBLE TORQUE WRENCH, XIA IS2217XLP MFG BY: STRYKER SPINE 2 PEARL COURT ALLENDALE NJ 07401-1677 The Xia 3/Mantis Redux/Specialty Audible Torque Wrenches are used to final tighten the blocker of the Xia 3/ Mantis Redux Polyaxial Screws, Hooks, Monoaxial Screws, Rod to Rod Connectors and Off-set connectors. The Torque Wrench is a t-handle deflection beam torque wrench with a 5 mm hex tip. The instrument allows the controlled application of 12 Nm torque via a visual indicator on the bell shaped cylinder just below the handle. The Specialty Audible Torque Wrench has an extra feature where it creates and audible "click" when the desired final tightening torque is reached.

Reason

Multiple complaints have reported the hex tip of the torque wrench fracturing during final tightening.

Code Info

Class I device, 510(K) exempt Catalog Number: IS2217XLP Lot Number: 2576, 2756

Distribution

Worldwide Distribution - USA (nationwide) and internationally to Australia, Canada, France, Germany, Netherlands, Japan, Spain, Denmark, Switzerland, and the United Kingdom.

Quantity

28 units