FDA Enforcement Class II Terminated

ORTHOFIX Catalogue Number: ref 1-1100201,DRILL BIT Diameter .4,8 MM L.240 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856068

Recall: Z-0381-2020 · Reported November 20, 2019

Enforcement

Recall Number
Z-0381-2020
Event ID
84004
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Orthofix Srl
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 20, 2019
Initiation Date
September 27, 2018
Classification Date
November 14, 2019
Termination Date
February 1, 2022
Address
Via Delle Nazioni 9 O 73, N/A, Veronaverona, N/A, N/A, Italy

Description

ORTHOFIX Catalogue Number: ref 1-1100201,DRILL BIT Diameter .4,8 MM L.240 MM TIN COATED - QUICK CONNECT, RX ONLY, Non Sterile, UDI: (01)18033509856068

Reason

As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.

Code Info

Lot # B1205238, B1205760, B1213842, 289814, 300239, 289814, 273343

Distribution

US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel

Quantity

N/A