FDA Enforcement Class II Ongoing

Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.

Recall: Z-0379-2018 · Reported February 7, 2018

Enforcement

Recall Number
Z-0379-2018
Event ID
78885
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Carl Zeiss Metrology Inc
Voluntary / Mandated
FDA Mandated
Initial Notification
Letter
Report Date
February 7, 2018
Initiation Date
November 14, 2017
Classification Date
January 26, 2018
Address
6250 Sycamore Ln N, Maple Grove, MN, 55369-6310, United States

Description

Metrotom 800 (130kV CT scanner) Industrial CT scanner for measuring and inspecting complete components made of plastic or light metal.

Reason

Due to a software error, it was observed that when the user closes the access door, the system will resume its scan cycle automatically, instead of the user initiating x-ray generation from the control panel.

Code Info

Serial Numbers: 152806 157499 157683 181193 181364 182887 186654 186767 186797 192099 192182 192351 197342

Distribution

Worldwide Distribution - US, including the states of : WI, NC, GA, IL, AZ, FL, CA, NJ, and territory of PR; and the country of Canada.

Quantity

13 US