FDA Enforcement Class III Terminated

Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .

Recall: Z-0379-2013 · Reported November 28, 2012

Enforcement

Recall Number
Z-0379-2013
Event ID
62987
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Microgenics Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 28, 2012
Initiation Date
August 20, 2012
Classification Date
November 19, 2012
Termination Date
March 4, 2013
Address
44660 Osgood Rd, N/A, Fremont, CA, 94539-6410, United States

Description

Thermo Scientific MAS Omni IMMUNE and Omni IMMUNE PRO; Liquid Assayed Integrated Immunoassay Control. Intended for use as an assayed control for monitoring assay conditions in many clinical laboratory determinations. Assay values are provided for the specific systems listed. The user can compare observations with expected ranges as a means of assuring consistent performance or reagent and instrument. .

Reason

In house testing confirmed that CK-MS and Insulin claimed in both Omni IMMUNE and Omni IMMUNE PRO products are displaying reduced stability.

Code Info

MAS Onmi IMMUNE: Catalog Number/lot number/ expiration date: OIM-101 - OIM14031 - 31 MAR 2014; OIM-202 - OIM14032 - 31 MAR 2014; OIM-303 - OIM14033 - 31 MAR 2014; OIM-SP - OIM1403S - 31 MAR 2014. MAS Omni IMMUNE PRO: OPRO-101 - OPRO14031 - 31 MAR 2014; OPRO-201 - OPRO14032 - 31 MAR 2014' OPRO-303 - OPRO14033 - 31 MAR 2014; OPRO-SP - OPRO1403S - 31 MAR 2014.

Distribution

Worldwide Distribution--USA (nationwide) and the countries of Germany, Canada and Australia.

Quantity

2076 total in US , 596 total Foreign