CODMAN Disposable Perforator 9mm. Cranial Perforator.
Enforcement
- Recall Number
- Z-0374-2026
- Event ID
- 97623
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Integra LifeSciences Corp. (NeuroSciences)
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 12, 2025
- Initiation Date
- September 26, 2025
- Classification Date
- November 3, 2025
- Address
- 1100 Campus Rd, N/A, Princeton, NJ, 08540-6650, United States
Description
CODMAN Disposable Perforator 9mm. Cranial Perforator.
Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.
Model No. 261223; UDI: 10381780513605; Lot No. 5828553, 6264478, 6264481, 6264482, 7230672, 7230673, 7230674, 7230675, 7261971, 7268008, 7268009, 7268010, 7268011, 7268012, 7268013, 7268014, 7268015, 7268016, 7268017, 7268018, 7268019, 7282954, 7282955, 7293422, 7293423, 7293424, 7293425, 7293426, 7293427, 7293428, 7300667, 7300668, 7300669, 7300670, 7300672, 7300673, 7300674, 7371082, 7371083, 7379652, 7379922, 7379923, 7379924, 7379926, 7379928, 7379660, 7379925, 7379927, 7450559, 7450560, 7450561, 7450562, 7450563, 7450564, 7450565, 7450566, 7451273, 7451274, 7451275, 7451276, 7451277, 7451278, 7451279; Expiration Date: 2024-03-31 to 2029-07-31.
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Barbados, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czechia, Ecuador, France, French Guiana, Germany, Greece, India, Ireland, Italy, Japan, Jordan, Korea, Lithuania, Mexico, Morrocco, Netherlands, New Zealand, Palestine, State of, Philippines, Portugal, Puerto Rico, Qatar, Republic of North Macedonia, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan , Thailand, Trinidad and Tobago, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Vietnam.
5,109 units