ORTHOFIX Catalogue Number: ref 11007, drill bit diameter 4.8 MM LENGTH 280 MM, RX ONLY, Non Sterile, UDI: (01)18032568031959
Recall: Z-0373-2020
·
Reported November 20, 2019
Enforcement
- Recall Number
- Z-0373-2020
- Event ID
- 84004
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Orthofix Srl
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 20, 2019
- Initiation Date
- September 27, 2018
- Classification Date
- November 14, 2019
- Termination Date
- February 1, 2022
- Address
- Via Delle Nazioni 9 O 73, N/A, Veronaverona, N/A, N/A, Italy
Description
ORTHOFIX Catalogue Number: ref 11007, drill bit diameter 4.8 MM LENGTH 280 MM, RX ONLY, Non Sterile, UDI: (01)18032568031959
Reason
As a result of complaint investigations it was determined that batches of drill bits have a geometrical detail that does not correspond to the intended design. This could result in lower cutting performance.
Code Info
Lot # B1200908
Distribution
US: AZ, CA,FL,GA,HI,IA,IL,IN,KY,LA,MD,MI,MO,MS,NC,NY,OH,OK,OR,PA,SC,TN,TX,UT,VA,WA, WI, WV, Puerto Rico OUS: Kuwait, Australia, Saudi Arabia, United Arab Emirates, Ireland, Japan, Spain, Poland, Switzerland, Singapore, Peru, Brazil, Germany, Great Britain, France, Italy, Australia, South Africa, Belgium, Chile, The Netherlands, Lithuania, Sri Lanka, Denmark, Israel
Quantity
N/A