FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
Enforcement
- Recall Number
- Z-0368-2023
- Event ID
- 90995
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Full Vision Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 14, 2022
- Initiation Date
- August 31, 2022
- Classification Date
- December 2, 2022
- Address
- 3017 Full Vision Dr, N/A, Newton, KS, 67114-9750, United States
Description
FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST1 110V, Model #317-07926GE
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
UDI/DI 00860176000668, Serial Numbers: GEDC-6632, GEDC-6633, GEDC-6634, GEDC-6635, GEDC-6636, GEDC-6637, GEDC-6638, GEDC-6639, GEDC-6640, GEDC-6641, GEDC-6642, GEDC-6643, GEDC-6644, GEDC-6645, GEDC-6646, GEDC-6647, GEDC-6648, GEDC-6649, GEDC-6650, GEDC-6651, GEDC-6652, GEDC-6653, GEDC-6654, GEDC-6655, GEDC-6656, GEDC-6657, GEDC-6658, GEDC-6659, GEDC-6660, GEDC-6661, GEDC-6662, GEDC-6663, GEDC-6664, GEDC-6665, GEDC-6666, GEDC-6667, GEDC-6668, GEDC-6669, GEDC-6670, GEDC-6671, GEDC-6672, GEDC-6673, GEDC-6674, GEDC-6675, GEDC-6676, GEDC-6677, GEDC-6678, GEDC-6679, GEDC-6680, GEDC-6681, GEDC-6755, GEDC-6756, GEDC-6757, GEDC-6758, GEDC-6759, GEDC-6760, GEDC-6761, GEDC-6762, GEDC-6763, GEDC-6764, GEDC-6839, GEDC-6840, GEDC-6841, GEDC-6842, GEDC-6843, GEDC-6844, GEDC-6845, GEDC-6846, GEDC-6847
Wi, GA, NY, Mexico
69 units