FDA Enforcement Class II Ongoing

Ventlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN4100MCP; 2. V-Care Infant Resuscitator REF #s: VN3100OB, VN3100MB-2, VN3100MB-M01, VNW3100MB-2, VN3100MBP-3M, VN3100MB, VN3100MB-D1, VN3100MB-M00-0, VN3100MB-M1, VN3100OBP, VN3100MB-B, VN3100MB-I, VN3100MBP-000, VN3100MBP-2, VN3100MB-PW2, VN3100MBP, VN3100MBP-D1, VN3100MBPW-D2, VN3100MBP-F, VN3100MXK-2, VN3100MX-F, VN3100MC-2, VN3100MC, VN3100MC-B, VN3100MC-D1, VN3100MC-2U, VN3106MC-M1-2V, VN3104MCFB- M01, VN3106MC-D2V, VN3106MC-M1V, VN3100MCP-2, VN3100MCP, VN3106MCPF- M1V, VN3100MBF; 3. V-Care Pediatric Resuscitator REF #s: VN2100MB, VN2102MB, VN2102MBF, VN2102OBP, VN2100MBF, VN2100MBP, VN2100MB-PW, VN2102MX, VN2102MX-F, VN2100MX-M123, VN2100MC, VN2102MC, VN2102MC-B, VN2102MCP, VN2100MCP; 4. V-Care Small Adult Resuscitator REF #s: VN5000MB, VN5002MB, VN5100MB-F, VN5000MBPF, VN5000MBPF-I, VN5000MX, VN5100MX-F, VN5000MX-N, VN5000MXP, VN5000MXP-K, VN5000MC, VN5102MC, VN5100MCP, VN5000MC-N.

Recall: Z-0367-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0367-2024
Event ID
93237
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SunMed Holdings, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 29, 2023
Initiation Date
October 11, 2023
Classification Date
November 21, 2023
Address
2710 Northridge Dr Nw Ste A, N/A, Grand Rapids, MI, 49544-9112, United States

Description

Ventlab, LLC V-Care manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. V-Care Child Resuscitator REF #s: VN4100MB, VN4100MCFB-3, VN4100MCP; 2. V-Care Infant Resuscitator REF #s: VN3100OB, VN3100MB-2, VN3100MB-M01, VNW3100MB-2, VN3100MBP-3M, VN3100MB, VN3100MB-D1, VN3100MB-M00-0, VN3100MB-M1, VN3100OBP, VN3100MB-B, VN3100MB-I, VN3100MBP-000, VN3100MBP-2, VN3100MB-PW2, VN3100MBP, VN3100MBP-D1, VN3100MBPW-D2, VN3100MBP-F, VN3100MXK-2, VN3100MX-F, VN3100MC-2, VN3100MC, VN3100MC-B, VN3100MC-D1, VN3100MC-2U, VN3106MC-M1-2V, VN3104MCFB- M01, VN3106MC-D2V, VN3106MC-M1V, VN3100MCP-2, VN3100MCP, VN3106MCPF- M1V, VN3100MBF; 3. V-Care Pediatric Resuscitator REF #s: VN2100MB, VN2102MB, VN2102MBF, VN2102OBP, VN2100MBF, VN2100MBP, VN2100MB-PW, VN2102MX, VN2102MX-F, VN2100MX-M123, VN2100MC, VN2102MC, VN2102MC-B, VN2102MCP, VN2100MCP; 4. V-Care Small Adult Resuscitator REF #s: VN5000MB, VN5002MB, VN5100MB-F, VN5000MBPF, VN5000MBPF-I, VN5000MX, VN5100MX-F, VN5000MX-N, VN5000MXP, VN5000MXP-K, VN5000MC, VN5102MC, VN5100MCP, VN5000MC-N.

Reason

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Code Info

ITEM # VN2100MB, UDI-DI: 10889483095379, (Case UDI: 30889483095373); ITEM # VN2100MBF, UDI-DI: 10889483092774, (Case UDI: 30889483092778); ITEM # VN2100MBP, UDI-DI: 10889483108260, (Case UDI: 30889483108264); ITEM # VN2100MB-PW, UDI-DI: 10889483108253, (Case UDI: 30889483108257); ITEM # VN2100MC, UDI-DI: 10889483108277, (Case UDI: 30889483108271); ITEM # VN2100MCP, UDI-DI: 10889483094525, (Case UDI: 30889483094529); ITEM # VN2100MX-M123, UDI-DI: 10889483098936, (Case UDI: 30889483098930); ITEM # VN2102MB, UDI-DI: 10889483105542, (Case UDI: 30889483105546); ITEM # VN2102MBF, UDI-DI: 10889483110966, (Case UDI: 30889483110960); ITEM # VN2102MC, UDI-DI: 10889483111994, (Case UDI: 30889483111998); ITEM # VN2102MC-B, UDI-DI: 10889483107393, (Case UDI: 30889483107397); ITEM # VN2102MCP, UDI-DI: 10889483088470, (Case UDI: 30889483088474); ITEM # VN2102MX, UDI-DI: 10889483095751, (Case UDI: 30889483095755); ITEM # VN2102MX-F, UDI-DI: 10889483105863, (Case UDI: 30889483105867); ITEM # VN2102OBP, UDI-DI: 10889483108895, (Case UDI: 30889483108899); ITEM # VN3100MB, UDI-DI: 10889483096994, (Case UDI: 30889483096998); ITEM # VN3100MB-2, UDI-DI: 10889483105900, (Case UDI: 30889483105904); ITEM # VN3100MB-B, UDI-DI: 10889483113585, (Case UDI: 30889483113589); ITEM # VN3100MB-D1, UDI-DI: 10889483105566, (Case UDI: 30889483105560); ITEM # VN3100MBF, UDI-DI: 10889483100622, (Case UDI: 30889483100626); ITEM # VN3100MB-I, UDI-DI: 10889483109380, (Case UDI: 30889483109384); ITEM # VN3100MB-M00- 0, UDI-DI: 10889483095980, (Case UDI: 30889483095984); ITEM # VN3100MB-M01, UDI-DI: 10889483110386, (Case UDI: 30889483110380); ITEM # VN3100MB-M1, UDI-DI: 10889483102763, (Case UDI: 30889483102767); ITEM # VN3100MBP, UDI-DI: 10889483100622, (Case UDI: 30889483100626); ITEM # VN3100MBP-000, UDI-DI: 10889483110393, (Case UDI: 30889483110397); ITEM # VN3100MBP-2, UDI-DI: 10889483104262, (Case UDI: 30889483104266); ITEM # VN3100MBP-3M, UDI-DI: 10889483100639, (Case UDI: 30889483100633); ITEM # VN3100MBP-D1, UDI-DI: 10889483110751, (Case UDI: 30889483110755); ITEM # VN3100MBP-F, UDI-DI: 10889483102770, (Case UDI: 30889483102774); ITEM # VN3100MB-PW2, UDI-DI: 10889483110744, (Case UDI: 30889483110748); ITEM # VN3100MBPW-D2, UDI-DI: 10889483111192, (Case UDI: 30889483111196); ITEM # VN3100MC, UDI-DI: 10889483100646, (Case UDI: 30889483100640); ITEM # VN3100MC-2, UDI-DI: 10889483111208, (Case UDI: 30889483111202); ITEM # VN3100MC-2U, UDI-DI: 10889483112113, (Case UDI: 30889483112117); ITEM # VN3100MC-B, UDI-DI: 10889483095768, (Case UDI: 30889483095762); ITEM # VN3100MC-D1, UDI-DI: 10889483112021, (Case UDI: 30889483112025); ITEM # VN3100MCP, UDI-DI: 10889483110768, (Case UDI: 30889483110762); ITEM # VN3100MCP-2, UDI-DI: 10889483113622, (Case UDI: 30889483113626); ITEM # VN3100MX-F, UDI-DI: 10889483112038, (Case UDI: 30889483112032); ITEM # VN3100MXK-2, UDI-DI: 10889483111215, (Case UDI: 30889483111219); ITEM # VN3100OB, UDI-DI: 10889483099964, (Case UDI: 30889483099968); ITEM # VN3100OBP, UDI-DI: 10889483105924, (Case UDI: 30889483105928); ITEM # VN3104MCFB- M01, UDI-DI: 10889483110775, (Case UDI: 30889483110779); ITEM # VN3106MC-D2V, UDI-DI: 10889483107539, (Case UDI: 30889483107533); ITEM # VN3106MC-M1- 2V, UDI-DI: 10889483095997, (Case UDI: 30889483095991); ITEM # VN3106MC-M1V, UDI-DI: 10889483097991, (Case UDI: 30889483097995); ITEM # VN3106MCPF- M1V, UDI-DI: 10889483099001, (Case UDI: 30889483099005); ITEM # VN4100MB, UDI-DI: 10889483111338, (Case UDI: 30889483111332); ITEM # VN4100MCFB-3, UDI-DI: 10889483111239, (Case UDI: 30889483111233); ITEM # VN4100MCP, UDI-DI: 10889483098042, (Case UDI: 30889483098046); ITEM # VN5000MB, UDI-DI: 10889483112052, (Case UDI: 30889483112056); ITEM # VN5000MBPF, UDI-DI: 10889483112229, (Case UDI: 30889483112223); ITEM # VN5000MBPF-I, UDI-DI: 10889483113646, (Case UDI: 30889483113640); ITEM # VN5000MC, UDI-DI: 10889483095393, (Case UDI: 30889483095397); ITEM # VN5000MC-N, UDI-DI: 10889483108352, (Case UDI: 30889483108356); ITEM # VN5000MX, UDI-DI: 10889483107782, (Case UDI: 30889483107786); ITEM # VN5000MX-N, UDI-DI: 10889483107553, (Case UDI: 30889483107557); ITEM # VN5000MXP, UDI-DI: 10889483089828, (Case UDI: 30889483089822); ITEM # VN5000MXP-K, UDI-DI: 10889483110409, (Case UDI: 30889483110403); ITEM # VN5002MB, UDI-DI: 10889483100660, (Case UDI: 30889483100664); ITEM # VN5100MB-F, UDI-DI: 10889483107560, (Case UDI: 30889483107564); ITEM # VN5100MCP, UDI-DI: 10889483098554, (Case UDI: 30889483098558); ITEM # VN5100MX-F, UDI-DI: 10889483095782, (Case UDI: 30889483095786); ITEM # VN5102MC, UDI-DI: 10889483109403, (Case UDI: 30889483109407); ITEM # VNW3100MB-2, UDI-DI: 10889483100677, (Case UDI: 30889483100671). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.

Distribution

US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.