FDA Enforcement Class II Ongoing

Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflation Bag Reservoir, Manometer, PEEP Valve, and 40cm Pop-Off REF # PRO-1927.

Recall: Z-0365-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0365-2024
Event ID
93237
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SunMed Holdings, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 29, 2023
Initiation Date
October 11, 2023
Classification Date
November 21, 2023
Address
2710 Northridge Dr Nw Ste A, N/A, Grand Rapids, MI, 49544-9112, United States

Description

Ventlab, LLC SafeSpot manual resuscitators with integrated manometer, multiple accessory configurations to include: 1. SafeSpot Infant Resuscitator REF #s: SS3200OBPC-T, SS3204MB-D12, SS3200MB-2, SS3200MB-D1, SS3200OBP, SS3200MB-MMC, SS3200MBP-M00, SS3200MB-PW, SS3200MBP-2I, SS3200MBF-2, SS3204MBF-2, SS3200MB- PWS10, SS3206MCP-S1, SS3204MB, SS3204MC-P2, SS3200MB, SS3200MBP-2, SS3200MBF, SS3200MB-I, SS3200MBP, SS3200OB, SS3200MC, SS3200MC-P2; 2. SafeSpot Safe Spot Bag, D1 Mask with Inflation Bag Reservoir, Manometer, PEEP Valve, and 40cm Pop-Off REF # PRO-1927.

Reason

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Code Info

UDI-DI: 10889483104107, (Case UDI: 30889483104105, Item # PRO-1927); UDI-DI: 10889483102954, (Case UDI: 30889483102958, Item # SS3200MB); UDI-DI: 10889483092590, (Case UDI: 30889483092594, Item # SS3200MB-2); UDI-DI: 10889483104330, (Case UDI: 30889483104334, Item #SS3200MB-D1); UDI-DI: 10889483172322, (Case UDI: 30889483172326, Item # SS3200MBF); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3200MBF-2); UDI-DI: 10889483104545, (Case UDI: 30889483104549, Item # SS3200MB-I); UDI-DI: 10889483105771, (Case UDI: 30889483105775, Item # SS3200MB-MMC); UDI-DI: 10889483086919, (Case UDI: 30889483086913, Item # SS3200MBP); UDI-DI: 10889483104903, (Case UDI: 30889483104907, Item # SS3200MBP-2); UDI-DI: 10889483107324, (Case UDI: 30889483107328, Item # SS3200MBP-2I); UDI-DI: 10889483103777, (Case UDI: 30889483103771, Item # SS3200MBP-M00); UDI-DI: 10889483103111, (Case UDI: 30889483103115, Item # SS3200MB-PW); UDI-DI: 10889483086902, (Case UDI: 30889483086906, Item # SS3200MB- PWS10); UDI-DI: 10889483103784, (Case UDI: 30889483103788, Item # SS3200MC); UDI-DI: 10889483103128, (Case UDI: 30889483103122, Item # SS3200MC-P2); UDI-DI: 10889483107904, (Case UDI: 30889483107908, Item # SS3200OB); UDI-DI: 10889483086926, (Case UDI: 30889483086920, Item # SS3200OBP); UDI-DI: 10889483231142, (Case UDI: 30889483231146, Item # SS3200OBPC-T); UDI-DI: 10889483107331, (Case UDI: 30889483107335, Item # SS3204MB); UDI-DI: 10889483158913, (Case UDI: 30889483158917, Item # SS3204MB-D12); UDI-DI: 10889483104354, (Case UDI: 30889483104358, Item # SS3204MBF-2); UDI-DI: 10889483160497, (Case UDI: 30889483160491, Item # SS3204MC-P2); UDI-DI: 10889483111734, (Case UDI: 30889483111738, Item # SS3206MCP-S1. All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.

Distribution

US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.

Quantity

N/A