FDA Enforcement Class II Terminated

Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050

Recall: Z-0365-2016 · Reported December 9, 2015

Enforcement

Recall Number
Z-0365-2016
Event ID
69090
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 9, 2015
Initiation Date
August 19, 2014
Classification Date
December 3, 2015
Termination Date
June 14, 2016
Address
3000 Minuteman Rd, Andover, MA, 01810-1032, United States

Description

Patient Support Material Number: 9890 010 87431, used with:EasyDiagnost ; Model: 706050

Reason

The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.

Code Info

All patient support for stitching are affected.

Distribution

Distributed in the states of MN, KY, MA, VA, TX, NC, HI, FL, AZ, CT, and WY and the countries of Argentina, Australia, Austria, Canada, Denmark, Germany, Netherlands, Russia, Saudi Arabia, Singapore, Spain, and Switzerland.

Quantity

260 units, not all units have the patient support accessory