Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6, AF5140MB-F, AF1140MB-T, AF1140MB-PW, AF5140MB-PW, AF1140MB-I, AF5140MB-I, AF5140MBN, AF1140MBS, AF5140MBS, AF1140MB-X, AF5140MB-X, AF1140MB-P, AF5140MBP, AF1140MB-K, AF5140MBS-I,AF5140MBPF-G10, AF5140MBP-G10, AF1140MBP-T, AF5140MBP-I, AF1140MBP-K, AF1140MBC-N, AF1140MBP-2, AF1140MC, AF5140MC, AF1140MCP, AF5140MC-P, AF1100MB, AF5100MB, AF5100MC, AF1140MC-M5, AF1140MB-TM5, AF1140MBP-TM5, AF1100MC, AF1140OB, AF1140MBP-M5, AFW1140MB,AF1140MBS-P, AF1000MBN-20, AF5102MBP, AF1100MB-N, AF1100MC-P, AF1140MB, AF1140MB-E, AF1140MB-N, AF1140MB-R, AF1140MBS-K, AF1140MBSP-I, AF1142MB, AF1100MBS, AF1100MB-MMC, AF1100MB-PWG1, AF1100MBP; 2. AirFlow Child Resuscitator REF #s: AF4100MCP, AF4140MB, AF4140MB-I, AF4140MBP, AF4140MB-PI, AF4100MB, AF4100MBP, AF4100OB, AF4140MB-D23, AF4140OBP, AF4140MCP-D3, AF4140MC; 3. AirFlow Infant Resuscitator REF #s: AF3140OB, AF3144MB, AF3104MB-PW, AF3100MBP- M0D1, AF3100MBP- M012, AF3140MB-M000, AF3140MB-D2, AF3140MB-D1, AF3140MB-S1, AF3140MB-2B, AF3140MB-I, AF3140MB-K, AF3100MB-PW, AF3100MB-PTS1, AF3100MB-M0- D1, AF3100OC, AF3100MC-2, AF3100MC, AF3106MC-2, AF3140MCP-K, AF3140MC-2B, AF3100MCP-2, AF3140MC- M0D12, AF3144MCP- M1M00, AF3100OB, AF3140MB, AF3140MB-M0, AF3140MB-2, AF3140MB-M1, AF3140MB-PW, AF3140MB-P, AF3140MB-D12I, AF3140MBP-2, AF3140MC, AF3140MCP, AF3140MCPF, AF3100MB, AF3100MB-M0, AF3100MB-2, AF3100MB-M1, AF3100MB-X, AF3100MBP, AF3100MB-M01, AF3100MBP-M01, AF3100MBP-2, AF3100MBP-D12, AF3100MCP; 4. AirFlow Pediatric Resuscitator REF #s: AF2100OB, AF2100MB-3, AF2102OB, AF2100MB-D5, AF2140OB, AF2102MBP-2, AF2102MBPW-D4, AF2102MBP-M3, AF2140MB-2, AFW2142OB, AF2140MB-D4, AF2140MB-D43, AF2140MB-P2, AF2144MB, AFW2142MB-2, AF2142MB-D3, AF2242MBP, AF2142MBS-I, AF2142MBP- 224OT, AF2140MBPW- D23, AF2140MB-I, AF2140MB-K, AF2140MB-S, AF2140MBP-K, AF2100MB-PW, AF2100MB-MMC, AF2102MBP, AF2100MB- PWTS4, AF2140MCPF, AF2100MC, AF2102MC, AF2142MC-M01, AF2140MC, AF2142MC-D34, AF2140MCP-D4, AF2140MC-PT, AF2140MCP-K, AF2142MCP, AF2100MCP, AF2102MB, AF2102MCP, AF2102MC-PT, AF2142OB, AF2142OBNS, AF2142MB, AF2142MB-3, AF2142MB-2, AF2142MB-M5E, AF2142MB-PW, AF2142MBP, AF2142MB-K, AF2142MBP-2, AF2142MC-3, AF2142MCP-D4, AF2140MB, AF2140MB-PW, AF2140MBP, AF1142OB, AF2100MB; 5. AirFlow Small Adult Resuscitator REF #s: AF5140MC-D6, AF5140MB-M5, AF5100MB-D5, AF5142MBP-M5, AF5140MBPW-D5, AF5140MBPWTD5, AF5140MBP-D5, AF5140MBP-D6, AF5140MBPW- M5, AF5140MBP-M5, AFN5140MB, AF5140MB-OT, AF5140MB-T, AFD5140MB-T, AF5140MB-K, AF5140MBS-R, AF5140MBS-P, AF5140MB-PT, AF5140MB-ST, AF5140MCP-K, AF5100MB-PW, AF5102MB-N, AF5140MBS-M5. 6. Ventlab, LLC Resuscitation Kit REF # MC1000.

Recall: Z-0361-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0361-2024
Event ID
93237
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
SunMed Holdings, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 29, 2023
Initiation Date
October 11, 2023
Classification Date
November 21, 2023
Address
2710 Northridge Dr Nw Ste A, N/A, Grand Rapids, MI, 49544-9112, United States

Description

Ventlab, LLC AirFlow manual resuscitator devices with integrated manometer, multiple accessory configurations to include: 1. AirFlow Adult Resuscitator REF #s: AF1140MB-M5, AF1140MB-D5, AF5142OB, AF5142MB, AF5142MB-D235, AF1142MBP, AF5142MBP, AF5140MB, AF5140MB-2, AF5140MB-D6, AF5140MB-F, AF1140MB-T, AF1140MB-PW, AF5140MB-PW, AF1140MB-I, AF5140MB-I, AF5140MBN, AF1140MBS, AF5140MBS, AF1140MB-X, AF5140MB-X, AF1140MB-P, AF5140MBP, AF1140MB-K, AF5140MBS-I,AF5140MBPF-G10, AF5140MBP-G10, AF1140MBP-T, AF5140MBP-I, AF1140MBP-K, AF1140MBC-N, AF1140MBP-2, AF1140MC, AF5140MC, AF1140MCP, AF5140MC-P, AF1100MB, AF5100MB, AF5100MC, AF1140MC-M5, AF1140MB-TM5, AF1140MBP-TM5, AF1100MC, AF1140OB, AF1140MBP-M5, AFW1140MB,AF1140MBS-P, AF1000MBN-20, AF5102MBP, AF1100MB-N, AF1100MC-P, AF1140MB, AF1140MB-E, AF1140MB-N, AF1140MB-R, AF1140MBS-K, AF1140MBSP-I, AF1142MB, AF1100MBS, AF1100MB-MMC, AF1100MB-PWG1, AF1100MBP; 2. AirFlow Child Resuscitator REF #s: AF4100MCP, AF4140MB, AF4140MB-I, AF4140MBP, AF4140MB-PI, AF4100MB, AF4100MBP, AF4100OB, AF4140MB-D23, AF4140OBP, AF4140MCP-D3, AF4140MC; 3. AirFlow Infant Resuscitator REF #s: AF3140OB, AF3144MB, AF3104MB-PW, AF3100MBP- M0D1, AF3100MBP- M012, AF3140MB-M000, AF3140MB-D2, AF3140MB-D1, AF3140MB-S1, AF3140MB-2B, AF3140MB-I, AF3140MB-K, AF3100MB-PW, AF3100MB-PTS1, AF3100MB-M0- D1, AF3100OC, AF3100MC-2, AF3100MC, AF3106MC-2, AF3140MCP-K, AF3140MC-2B, AF3100MCP-2, AF3140MC- M0D12, AF3144MCP- M1M00, AF3100OB, AF3140MB, AF3140MB-M0, AF3140MB-2, AF3140MB-M1, AF3140MB-PW, AF3140MB-P, AF3140MB-D12I, AF3140MBP-2, AF3140MC, AF3140MCP, AF3140MCPF, AF3100MB, AF3100MB-M0, AF3100MB-2, AF3100MB-M1, AF3100MB-X, AF3100MBP, AF3100MB-M01, AF3100MBP-M01, AF3100MBP-2, AF3100MBP-D12, AF3100MCP; 4. AirFlow Pediatric Resuscitator REF #s: AF2100OB, AF2100MB-3, AF2102OB, AF2100MB-D5, AF2140OB, AF2102MBP-2, AF2102MBPW-D4, AF2102MBP-M3, AF2140MB-2, AFW2142OB, AF2140MB-D4, AF2140MB-D43, AF2140MB-P2, AF2144MB, AFW2142MB-2, AF2142MB-D3, AF2242MBP, AF2142MBS-I, AF2142MBP- 224OT, AF2140MBPW- D23, AF2140MB-I, AF2140MB-K, AF2140MB-S, AF2140MBP-K, AF2100MB-PW, AF2100MB-MMC, AF2102MBP, AF2100MB- PWTS4, AF2140MCPF, AF2100MC, AF2102MC, AF2142MC-M01, AF2140MC, AF2142MC-D34, AF2140MCP-D4, AF2140MC-PT, AF2140MCP-K, AF2142MCP, AF2100MCP, AF2102MB, AF2102MCP, AF2102MC-PT, AF2142OB, AF2142OBNS, AF2142MB, AF2142MB-3, AF2142MB-2, AF2142MB-M5E, AF2142MB-PW, AF2142MBP, AF2142MB-K, AF2142MBP-2, AF2142MC-3, AF2142MCP-D4, AF2140MB, AF2140MB-PW, AF2140MBP, AF1142OB, AF2100MB; 5. AirFlow Small Adult Resuscitator REF #s: AF5140MC-D6, AF5140MB-M5, AF5100MB-D5, AF5142MBP-M5, AF5140MBPW-D5, AF5140MBPWTD5, AF5140MBP-D5, AF5140MBP-D6, AF5140MBPW- M5, AF5140MBP-M5, AFN5140MB, AF5140MB-OT, AF5140MB-T, AFD5140MB-T, AF5140MB-K, AF5140MBS-R, AF5140MBS-P, AF5140MB-PT, AF5140MB-ST, AF5140MCP-K, AF5100MB-PW, AF5102MB-N, AF5140MBS-M5. 6. Ventlab, LLC Resuscitation Kit REF # MC1000.

Reason

A backwards leak present in the integrated manometer of the patient valve allows for CO2 rebreathing.

Code Info

UDI-DI: 10889483075814 (Item # AF1000MBN-20, Case UDI: 30889483075818); UDI-DI: 10889483076606 (Item # AF1100MB, Case UDI: 30889483076600); UDI-DI: 10889483076613 (Item # AF1100MB-MMC, Case UDI: 30889483076617); UDI-DI: 10889483075838 (Item # AF1100MB-N, Case UDI: 30889483075832); UDI-DI: 10889483088074 (Item # AF1100MBP, Case UDI: 30889483088078); UDI-DI: 10889483086209 (Item # AF1100MB-PWG1, Case UDI: 30889483086203); UDI-DI: 10889483086216 (Item # AF1100MBS, Case UDI: 30889483086210); UDI-DI: 10889483094365 (Item # AF1100MC, Case UDI: 30889483094369); UDI-DI: 10889483159217 (Item # AF1100MC-P, Case UDI: 30889483159211); UDI-DI: 10889483071809 (Item # AF1140MB, Case UDI: 30889483071803); UDI-DI: 10889483230626 (Item # AF1140MBC-N Case UDI: 30889483230620); UDI-DI: 10889483090718 (Item # AF1140MB-D5, Case UDI: 30889483090712); UDI-DI: 10889483106372 (Item # AF1140MB-E, Case UDI: 30889483106376); UDI-DI: 10889483095096 (Item # AF1140MB-I, Case UDI: 30889483095090); UDI-DI: 10889483083918 (Item # AF1140MB-K, Case UDI: 30889483083912); UDI-DI: 10889483162224 (Item # AF1140MB-M5, Case UDI: 30889483162228); UDI-DI: 10889483094372 (Item # AF1140MB-N, Case UDI: 30889483094376); UDI-DI: 10889483098370 (Item # AF1140MB-P, Case UDI: 30889483098374); UDI-DI: 10889483101469 (Item # AF1140MBP-2, Case UDI: 30889483101463); UDI-DI: 10889483106396 (Item # AF1140MBP-K, Case UDI: 30889483106390); UDI-DI: 10889483076637 (Item # AF1140MBP-M5, Case UDI: 30889483076631); UDI-DI: 10889483078143 (Item # AF1140MBP-T, Case UDI: 30889483078147); UDI-DI: 10889483162637 (Item # AF1140MBP-TM5, Case UDI: 30889483162631); UDI-DI: 10889483098387 (Item # AF1140MB-PW, Case UDI: 30889483098381); UDI-DI: 10889483074190 (Item # AF1140MB-R, Case UDI: 30889483074194); UDI-DI: 10889483095119 (Item # AF1140MBS, Case UDI: 30889483095113); UDI-DI: 10889483093320 (Item # AF1140MBS-K, Case UDI: 30889483093324); UDI-DI: 10889483076422 (Item # AF1140MBS-P, Case UDI: 30889483076426); UDI-DI: 10889483089088 (Item # AF1140MBSP-I, Case UDI: 30889483089082); UDI-DI: 10889483087213 (Item # AF1140MB-T, Case UDI: 30889483087217); UDI-DI: 10889483162439 (Item # AF1140MB-TM5, Case UDI: 30889483162433); UDI-DI: 10889483227589 (Item # AF1140MB-X, Case UDI: 30889483227583); UDI-DI: 10889483101476 (Item # AF1140MC, Case UDI: 30889483101470); UDI-DI: 10889483163146 (Item # AF1140MC-M5, Case UDI: 30889483163140); UDI-DI: 10889483097359 (Item # AF1140MCP, Case UDI: 30889483097353); UDI-DI: 10889483102121 (Item # AF1140OB, Case UDI: 30889483102125); UDI-DI: 10889483078150 (Item # AF1142MB, Case UDI: 30889483078154); UDI-DI: 10889483089255 (Item # AF1142MBP, Case UDI: 30889483089259); UDI-DI: 10889483089095 (Item # AF1142OB, Case UDI: 30889483089099); UDI-DI: 10889483096390 (Item # AF2100MB, Case UDI: 30889483096394); UDI-DI: 10889483076019 (Item # AF2100MB-3, Case UDI: 30889483076013); UDI-DI: 10889483171578 (Item # AF2100MB-D5, Case UDI: 30889483171572); UDI-DI: 10889483078181 (Item # AF2100MB-MMC, Case UDI: 30889483078185); UDI-DI: 10889483159231 (Item # AF2100MB-PW, Case UDI: 30889483159235); UDI-DI: 10889483102138 (Item # AF2100MB- PWTS4, Case UDI: 30889483102132); UDI-DI: 10889483090749 (Item # AF2100MC, Case UDI: 30889483090743); UDI-DI: 10889483101506 (Item # AF2100MCP, Case UDI: 30889483101500); UDI-DI: 10889483095706 (Item # AF2100OB, Case UDI: 30889483095700); UDI-DI: 10889483098394 (Item # AF2102MB, Case UDI: 30889483098398); UDI-DI: 10889483112977 (Item # AF2102MBP, Case UDI: 30889483112971); UDI-DI: 10889483076644 (Item # AF2102MBP-2, Case UDI: 30889483076648); UDI-DI: 10889483097304 (Item # AF2102MBP-M3, Case UDI: 30889483097308); UDI-DI: 10889483172285 (Item # AF2102MBPW-D4, Case UDI: 30889483172289); UDI-DI: 10889483158883 (Item # AF2102MC, Case UDI: 30889483158887); UDI-DI: 10889483078198 (Item # AF2102MCP, Case UDI: 30889483078192); UDI-DI: 10889483085769 (Item # AF2102MC-PT, Case UDI: 30889483085763); UDI-DI: 10889483077924 (Item # AF2102OB, Case UDI: 20889483077921); UDI-DI: 10889483078068 (Item # AF2140MB, Case UDI: 30889483078062); UDI-DI: 10889483081143 (Item # AF2140MB-2, Case UDI: 30889483081147); UDI-DI: 10889483076286 (Item # AF2140MB-D4, Case UDI: 30889483076280); UDI-DI: 10889483112984 (Item # AF2140MB-D43, Case UDI: 30889483112988); UDI-DI: 10889483077566 (Item # AF2140MB-I, Case UDI: 20889483077563); UDI-DI: 10889483081150 (Item # AF2140MB-K, Case UDI: 20889483081157); UDI-DI: 10889483088111 (Item # AF2140MBP, Case UDI: 30889483088115); UDI-DI: 10889483096413 (Item # AF2140MB-P2, Case UDI: 30889483096417); UDI-DI: 10889483083734 (Item # AF2140MBP-K, Case UDI: 30889483083738); UDI-DI: 10889483077573 (Item # AF2140MB-PW, Case UDI: 30889483077577); UDI-DI: 10889483159309 (Item # AF2140MBPW- D23, Case UDI: 30889483159303); UDI-DI: 10889483100424 (Item # AF2140MB-S, Case UDI: 30889483100428); UDI-DI: 10889483097311 (Item # AF2140MC, Case UDI: 20889483097318); UDI-DI: 10889483163344 (Item # AF2140MCP-D4, Case UDI: 30889483163348); UDI-DI: 10889483275788 (Item # AF2140MCPF, Case UDI: 30889483275782); UDI-DI: 10889483158821 (Item # AF2140MCP-K, Case UDI: 30889483158825); UDI-DI: 10889483159347 (Item # AF2140MC-PT, Case UDI: 30889483159341); UDI-DI: 10889483085776 (Item # AF2140OB, Case UDI: 30889483085770); UDI-DI: 10889483090237 (Item # AF2142MB , Case UDI: 30889483090231); UDI-DI: 10889483081167 (Item # AF2142MB-2, Case UDI: 30889483081161); UDI-DI: 10889483077931 (Item # AF2142MB-3, Case UDI: 30889483077935); UDI-DI: 10889483087756 (Item # AF2142MB-D3, Case UDI: 30889483087750); UDI-DI: 10889483088128 (Item # AF2142MB-K, Case UDI: 20889483088125); UDI-DI: 10889483102978 (Item # AF2142MBP, Case UDI: 30889483102972); UDI-DI: 10889483073315 (Item # AF2142MBP-2, Case UDI: 30889483073319); UDI-DI: 10889483097328 (Item # AF2142MBP- 224OT, Case UDI: 20889483097325); UDI-DI: 10889483102961 (Item # AF2142MB-PW, Case UDI: 20889483102968); UDI-DI: 10889483081174 (Item # AF2142MBS-I, Case UDI: 30889483081178); UDI-DI: 10889483081181 (Item # AF2142MC-3 , Case UDI: 30889483081185); UDI-DI: 10889483164136 (Item # AF2142MC-D34, Case UDI: 30889483164130); UDI-DI: 10889483168554 (Item # AF2142MC-M01, Case UDI: 30889483168558); UDI-DI: 10889483087800, (Item # AF2142MCP, Case UDI: 30889483087804); UDI-DI: 10889483167083 (Item # AF2142MCP-D4, Case UDI: 30889483167087); UDI-DI: 10889483089262 (Item # AF2142OB, Case UDI: 30889483089266); UDI-DI: 10889483226285 (Item # AF2142OBNS, Case UDI: 30889483226289); UDI-DI: 10889483078204 (Item # AF2144MB, Case UDI: 20889483078201); UDI-DI: 10889483555675 (Item # AF2142MB-M5E, Case UDI: 30889483555679); UDI-DI: 10889483077948 (Item # AF2242MBP, Case UDI: 20889483077945); UDI-DI: 10889483093542 (Item # AF3100MB, Case UDI: 30889483093546); UDI-DI: 10889483165171 (Item # AF3100MB-2, Case UDI: 30889483165175); UDI-DI: 10889483164556 (Item # AF3100MB-M0, Case UDI: 30889483164550); UDI-DI: 10889483085790 (Item # AF3100MB-M01, Case UDI: 20889483085797); UDI-DI: 10889483100431 (Item # AF3100MB-M0- D1, Case UDI: 30889483100435); UDI-DI: 10889483103432 (Item # AF3100MB-M1, Case UDI: 30889483103436); UDI-DI: 10889483089293 (Item # AF3100MBP, Case UDI: 30889483089297); UDI-DI: 10889483172254 (Item # AF3100MBP-2, Case UDI: 30889483172258); UDI-DI: 10889483159323 (Item # AF3100MBP-D12, Case UDI: 30889483159327); UDI-DI: 10889483083741 (Item # AF3100MBP-M01 Case UDI: 30889483083745 UDI-DI: 10889483158807 (Item # AF3100MBP- M012, Case UDI: 30889483158801); UDI-DI: 10889483165188 (Item # AF3100MBP- M0D1, Case UDI: 30889483165182); UDI-DI: 10889483087770 (Item # AF3100MB-PTS1, Case UDI: 30889483087774); UDI-DI: 10889483097335 (Item # AF3100MB-PW, Case UDI: 30889483097339); UDI-DI: 10889483228708 (Item # AF3100MB-X, Case UDI: 30889483228702); UDI-DI: 10889483081198 (Item # AF3100MC, Case UDI: 20889483081195); UDI-DI: 10889483077108 (Item # AF3100MC-2, Case UDI: 20889483077105); UDI-DI: 10889483089309 (Item # AF3100MCP, Case UDI: 30889483089303); UDI-DI: 10889483107447 (Item # AF3100MCP-2, Case UDI: 30889483107441); UDI-DI: 10889483075517 (Item # AF3100OB, Case UDI: 20889483075514); UDI-DI: 10889483097397 (Item # AF3100OC, Case UDI: 20889483097394); UDI-DI: 10889483172629 (Item # AF3104MB-PW, Case UDI: 30889483172623); UDI-DI: 10889483080023 (Item # AF3106MC-2, Case UDI: 30889483080027); UDI-DI: 10889483095713 (Item # AF3140MB, Case UDI: 30889483095717); UDI-DI: 10889483087220 (Item # AF3140MB-2, Case UDI: 30889483087224); UDI-DI: 10889483077115 (Item # AF3140MB-2B, Case UDI: 30889483077119); UDI-DI: 10889483078099 (Item # AF3140MB-D1, Case UDI: 30889483078093); UDI-DI: 10889483102985 (Item # AF3140MB-D12I, Case UDI: 30889483102989); UDI-DI: 10889483097342 (Item # AF3140MB-D2, Case UDI: 30889483097346); UDI-DI: 10889483087824 (Item # AF3140MB-I, Case UDI: 30889483087828); UDI-DI: 10889483103449 (Item # AF3140MB-K, Case UDI: 30889483103443); UDI-DI: 10889483077214 (Item # AF3140MB-M0, Case UDI: 30889483077218); UDI-DI: 10889483082683 (Item # AF3140MB-M000, Case UDI: 30889483082687); UDI-DI: 10889483102992 (Item # AF3140MB-M1, Case UDI: 30889483102996); UDI-DI: 10889483113011 (Item # AF3140MB-P, Case UDI: 30889483113015); UDI-DI: 10889483096444 (Item # AF3140MBP-2, Case UDI: 30889483096448); UDI-DI: 10889483159910 (Item # AF3140MB-PW, Case UDI: 30889483159914); UDI-DI: 10889483090299 (Item # AF3140MB-S1, Case UDI: 30889483090293); UDI-DI: 10889483079256 (Item # AF3140MC, Case UDI: 30889483079250); UDI-DI: 10889483077597 (Item # AF3140MC-2B, Case UDI: 30889483077591); UDI-DI: 10889483232088 (Item # AF3140MC- M0D12 UDI-DI: 10889483166963 (Item # AF3140MCP, Case UDI: 30889483166967); UDI-DI: 10889483275771 (Item # AF3140MCPF, Case UDI: 30889483275775); UDI-DI: 1088948315881 (Item # AF3140MCP-K, Case UDI: 30889483158818); UDI-DI: 10889483097410 (Item # AF3140OB, Case UDI: 30889483097414); UDI-DI: 10889483078105 (Item # AF3144MB, Case UDI: 30889483078109); UDI-DI: 10889483272626 (Item # AF3144MCP- M1M00, Case UDI: 30889483272620); UDI-DI: 10889483090305 (Item # AF4100MB, Case UDI: 30889483090309); UDI-DI: 10889483160121 (Item # AF4100MBP, Case UDI: 30889483160125); UDI-DI: 10889483182246 (Item # AF4100MCP, Case UDI: 30889483182240); UDI-DI: 10889483098417 (Item # AF4100OB, Case UDI: 30889483098411); UDI-DI: 10889483077627 (Item # AF4140MB, Case UDI: 30889483077621); UDI-DI: 10889483090510 (Item # AF4140MB-D23, Case UDI: 30889483090514); UDI-DI: 10889483093559 (Item # AF4140MB-I, Case UDI: 30889483093553); UDI-DI: 10889483099209 (Item # AF4140MBP, Case UDI: 30889483099203); UDI-DI: 10889483079287 (Item # AF4140MB-PI, Case UDI: 30889483079281); UDI-DI: 10889483087237 (Item # AF4140MC, Case UDI: 30889483087231); UDI-DI: 10889483163337 (Item # AF4140MCP-D3, Case UDI: 30889483163331); UDI-DI: 10889483090527 (Item # AF4140OBP, Case UDI: 30889483090521); UDI-DI: 10889483089323 (Item # AF5100MB, Case UDI: 30889483089327); UDI-DI: 10889483089330 (Item # AF5100MB-D5, Case UDI: 30889483089334); UDI-DI: 10889483159224 (Item # AF5100MB-PW, Case UDI: 30889483159228); UDI-DI: 10889483099346 (Item # AF5100MC, Case UDI: 30889483099340); UDI-DI: 10889483171929 (Item # AF5102MB-N, Case UDI: 30889483171923); UDI-DI: 10889483165164 (Item # AF5102MBP, Case UDI: 30889483165168); UDI-DI: 10889483103012 (Item # AF5140MB, Case UDI: 30889483103016); UDI-DI: 10889483079881 (Item # AF5140MB-2, Case UDI: 30889483079885); UDI-DI: 10889483089606 (Item # AF5140MB-D6, Case UDI: 30889483089600); UDI-DI: 10889483083758 (Item # AF5140MB-F, Case UDI: 30889483083752); UDI-DI: 10889483087251 (Item # AF5140MB-I, Case UDI: 30889483087255); UDI-DI: 10889483082720 (Item # AF5140MB-K, Case UDI: 30889483082724); UDI-DI: 10889483162408 (Item # AF5140MB-M5, Case UDI: 30889483162402); UDI-DI: 10889483090275 (Item # AF5140MBN, Case UDI: 30889483090279); UDI-DI: 10889483103029 (Item # AF5140MB-OT, Case UDI: 30889483103023); UDI-DI 10889483074725 (Item # AF5140MBP, Case UDI: 30889483074729 UDI-DI: 10889483098431 (Item # AF5140MBP-D5, Case UDI: 30889483098435); UDI-DI: 10889483104682 (Item # AF5140MBP-D6, Case UDI: 30889483104686); UDI-DI: 10889483275757 (Item # AF5140MBPF-G10, Case UDI: 30889483275751); UDI-DI: 10889483171967 (Item # AF5140MBP-G10, Case UDI: 30889483171961); UDI-DI: 10889483072233 (Item # AF5140MBP-I , Case UDI: 30889483072237); UDI-DI: 10889483163085 (Item # AF5140MBP-M5, Case UDI: 30889483163089); UDI-DI: 10889483089347 (Item # AF5140MB-PT, Case UDI: 30889483089341); UDI-DI: 10889483080061 (Item # AF5140MB-PW, Case UDI: 30889483080065); UDI-DI: 10889483099216 (Item # AF5140MBPW-D5, Case UDI: 30889483099210); UDI-DI: 10889483162514 (Item # AF5140MBPW- M5, Case UDI: 30889483162518); UDI-DI: 10889483079133 (Item # AF5140MBPWTD5, Case UDI: 30889483079137); UDI-DI: 10889483083772 (Item # AF5140MBS, Case UDI: 30889483083776); UDI-DI: 10889483077139 (Item # AF5140MBS-I, Case UDI: 30889483077133); UDI-DI: 10889483162231 (Item # AF5140MBS-M5, Case UDI: 30889483162235); UDI-DI: 10889483087268 (Item # AF5140MBS-P, Case UDI: 30889483087262); UDI-DI: 10889483083789 (Item # AF5140MBS-R, Case UDI: 30889483083783); UDI-DI: 10889483092682 (Item # AF5140MB-ST, Case UDI: 30889483092686); UDI-DI: 10889483083765 (Item # AF5140MB-T, Case UDI: 30889483083769); UDI-DI: 10889483227619 (Item # AF5140MB-X, Case UDI: 30889483227613); UDI-DI: 10889483090282 (Item # AF5140MC, Case UDI: 30889483090286); UDI-DI: 10889483228241 (Item # AF5140MC-D6, Case UDI: 30889483228245); UDI-DI: 10889483104200 (Item # AF5140MC-P, Case UDI: 30889483104204); UDI-DI: 10889483158838 (Item # AF5140MCP-K, Case UDI: 30889483158832); UDI-DI: 10889483089613 (Item # AF5142MB, Case UDI: 30889483089617); UDI-DI: 10889483227305 (Item # AF5142MB-D235, Case UDI: 30889483227309); UDI-DI: 10889483091838 (Item # AF5142MBP, Case UDI: 30889483091832); UDI-DI: 10889483163177 (Item # AF5142MBP-M5, Case UDI: 30889483163171); UDI-DI: 10889483080078 (Item # AF5142OB, Case UDI: 30889483080072); UDI-DI: 10889483085127 (Item # AFD5140MB-T, Case UDI: 30889483085121); UDI-DI: 10889483162781 (Item # AFN5140MB, Case UDI: 30889483162785); UDI-DI: 10889483099247 (Item # AFW1140MB, Case UDI: 30889483099241 UDI-DI: 10889483092705 (Item# AFW2142MB-2, Case UDI: 30889483092709); UDI-DI: 10889483078709 (Item # AFW2142OB, Case UDI: 30889483078703). All lots manufactured from 9/21/2018 to present. Latest expiration Date 10/11/2028.

Distribution

US Nationwide distribution including in the states of AR, AZ, CA¿CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MN, NC, ND, NH, NJ, NM, NV, NY, OH, SC, TN, TX, UT, VA, & WV.

Quantity