FDA Enforcement
Class II
Ongoing
Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.
Recall: Z-0361-2023
·
Reported December 14, 2022
Enforcement
- Recall Number
- Z-0361-2023
- Event ID
- 91087
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Cardinal Health
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 14, 2022
- Initiation Date
- November 1, 2022
- Classification Date
- December 2, 2022
- Address
- 777 West St, N/A, Mansfield, MA, 02048-1122, United States
Description
Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.
Reason
Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention
Code Info
UDI-DI: Each 10884521009585 Case 20884521009582 Lot Numbers: 2216806064, 2214452064, 2212401164 2210219764, 2209525364, 2208816464 2206737964, 2206727564
Distribution
Nationwide
Quantity
N/A