FDA Enforcement Class II Ongoing

Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.

Recall: Z-0361-2023 · Reported December 14, 2022

Enforcement

Recall Number
Z-0361-2023
Event ID
91087
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Cardinal Health
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 14, 2022
Initiation Date
November 1, 2022
Classification Date
December 2, 2022
Address
777 West St, N/A, Mansfield, MA, 02048-1122, United States

Description

Dover Closed Urethral Tray with Vinyl Catheter 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 1000 mL Drainage Bag Product Code: 3450LF Intended for urinary drainage from the bladder of a patient.

Reason

Occlusion (blockage) of the bubble inlet port of the urinary drainage bag which may prevent urine draining into the bag. If the bladder is not emptied within a clinically appropriate time, this may increase the risk of urinary retention

Code Info

UDI-DI: Each 10884521009585 Case 20884521009582 Lot Numbers: 2216806064, 2214452064, 2212401164 2210219764, 2209525364, 2208816464 2206737964, 2206727564

Distribution

Nationwide

Quantity

N/A