FDA Enforcement Class II Terminated

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Recall: Z-0360-2020 · Reported November 20, 2019

Enforcement

Recall Number
Z-0360-2020
Event ID
84135
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Canon Medical System, USA, INC.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 20, 2019
Initiation Date
February 8, 2019
Classification Date
November 12, 2019
Termination Date
June 7, 2022
Address
2441 Michelle Dr, N/A, Tustin, CA, 92780-7047, United States

Description

Infinix DP- i(Infinix-8000F) CAT-870B Catheterization Table (Titling) - Product Usage: The catheterization table is used with Infinix-8000F System (Interventional Angiography System), to position patient during diagnostic and interventional x-ray angiography.

Reason

Table tilts without operator involvement as a result of liquid spilling into tilt console triggering a short circuit.

Code Info

Model CAT-870B D053991 Serial No. Manufacture Date 99D08X2101 October 2008 99D0582039 August 2005 99D0762072 June 2007 99D0852090 May 2008 99D07Y2078 November 2007 99D0812081 January 2008 99D0582042 August 2005 99D0552033 May 2005 D551206 May 2005

Distribution

Worldwide distribution - US Nationwide distribution in the states of AL, CT, FL, IL, OR, PA, and TX; and country of Netherlands.

Quantity

9 systems