Philips Zenition 50, Model Number: 718096
Enforcement
- Recall Number
- Z-0352-2025
- Event ID
- 95579
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 20, 2024
- Initiation Date
- October 7, 2024
- Classification Date
- November 12, 2024
- Address
- Veenpluis 4-6, N/A, Best, N/A, N/A, Netherlands
Description
Philips Zenition 50, Model Number: 718096
Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
Model Number: 718096; UDI-DI: (01)00884838091535(21); Serial Numbers: 792, 1762, 1763, 1748, 236, 1310, 1311, 1764, 1078, 1759, 1760, 1761, 317, 1297, 75, 73, 939, 1517, 877, 1857, 50, 1286, 296, 580, 147, 477, 1681, 1288, 875, 1309, 466, 467, 1192, 1369, 386, 387, 52, 51, 1689, 1691, 1692, 1807, 4, 1690, 1693, 415, 1561, 1269, 432, 798, 799, 858, 860, 936, 944, 602, 676, 701, 702, 704, 705, 706, 293, 1895, 703, 859, 862, 863, 1010, 619, 650, 649, 674, 1806, 632, 613, 651, 750, 559, 1030, 886, 486, 487, 1356, 558, 560, 593, 1259, 1703, 1025, 1537, 139, 1206, 1207, 648, 1208, 1110, 372, 1392, 1607, 1248, 1687, 1688, 1686, 1168, 1700, 1232, 1256, 1550, 1068, 814, 1601, 1534, 1800, 1859, 1911, 1390, 975, 1321, 1251, 1673, 1793, 1794, 1795, 1796, 1942, 1249, 1250, 1252, 1253, 1338, 1340, 1342, 1532, 1533, 1684, 345, 400, 401, 402, 615, 617, 618, 768, 769, 770, 771, 772, 773, 774, 775, 620, 1405, 972, 1812, 1339, 1936, 1937, 1938, 1939, 596, 616, 111, 776, 428, 905, 1408, 1355, 1772, 1773, 678, 1313, 1831, 382, 904, 407, 187, 595, 1270, 1469, 546, 238, 479, 727, 728, 1854, 67, 578, 1801, 943, 223, 94, 1128, 1129, 248, 1016, 1211, 1788, 938, 1105, 1708, 1898, 497, 1672, 462, 1067, 1838, 435, 62, 63, 65, 1071, 810, 589, 590, 1073, 587, 1604, 380, 445, 140, 206, 146, 551, 1597, 91, 460, 47, 5, 92, 1069, 441, 459, 229, 24, 871, 586, 1225, 437, 1315, 143, 23, 1671, 1032, 622, 1538, 1033, 1634, 638, 259, 189, 1685, 191, 195, 1782, 689, 1092, 1224, 1755, 1015, 1057, 1351, 1832, 1833, 1667, 1555, 1187, 1527, 1528, 1214, 930, 1410, 1526, 866, 1220, 1529, 1221, 110, 1215, 1698, 929, 1011, 955, 1218, 1411, 1912, 1914, 1081, 1091, 654, 1352, 1769, 1535, 969, 200, 600, 1923, 1694, 1086, 793, 1654, 57, 512, 58, 608, 1231, 1683, 636, 971, 221, 1554, 1696, 1337, 1370, 1064, 362, 1599, 1357, 507, 594, 1054, 1810, 348, 688, 1605, 584, 430, 1877, 335, 336, 337, 338, 340, 341, 342, 343, 344, 367, 405, 76, 1205, 391, 1875, 1874, 1399, 1520, 1523, 1400, 1087, 1855, 516, 1768, 1201, 1846, 1185, 1562, 1525, 1867, 263, 1521, 923, 1975, 1870, 1976, 1422, 1853, 1198, 1659, 885, 1880, 1619, 1391, 783, 1582, 1883, 723, 1581, 1415, 1620, 79, 1296, 56, 720, 1341, 49, 1887, 876, 882, 888, 924, 935, 1402, 605, 606, 947, 683, 684, 1278, 1279, 1280, 1282, 1283, 1284, 1285, 1616, 1844, 1809, 1281, 207, 681, 682, 1088, 680, 1141, 1845, 320, 598, 270, 691, 788, 1567, 653, 1055, 542, 426, 427, 547, 524, 1090, 846, 456, 163, 205, 847, 505, 475, 476, 250, 1127, 1553, 1531, 1468, 1556, 1825, 1056, 1362, 1758, 1014, 1414, 968, 1602, 976, 977, 1372, 1603, 806, 1676, 1378, 1083, 1642, 1052, 1181, 1353, 1518, 1416, 951, 1780, 614, 949, 1701, 1702, 1989, 940, 510, 1822, 1821, 1066, 1239, 220, 898, 261, 494, 495, 1516, 1080, 87, 361, 357, 873, 899, 1519, 1826, 1147, 1097, 1682, 179, 180, 181, 1336, 1118, 766, 1074, 178, 1095, 188, 1034, 1035, 765, 1349, 767, 1199, 1200, 265, 1551, 421, 1036, 1814, 1749, 227, 228, 461, 1670, 1119, 256, 960, 961, 1076, 1404, 318, 142, 45, 46, 588, 519, 216, 359, 48, 96, 1360, 1358, 1359, 1652, 1900, 1365, 131, 1028, 126, 1649, 1367, 1019, 1020, 1102, 1178, 1180, 1216, 1217, 1219, 1272, 130, 132, 1375, 1542, 1543, 1547, 1600, 1661, 173, 20, 374, 540, 818, 928, 1544, 1515, 1179, 155, 1368, 1287, 533, 534, 535, 536, 537, 538, 747, 748, 1364, 667, 1202, 1024, 1879, 1934, 1665, 870, 1223, 133, 1924, 412, 1876, 1366, 1666, 499, 151, 1173, 1545, 473, 1021, 1022, 1273, 1017, 474, 819, 869, 1650, 129, 1171, 1271, 1733, 669, 1881, 1228, 1662, 1744, 1754, 1962, 1925, 666, 1664, 1317, 1318, 1169, 1125, 1226, 134, 668, 1653, 1277, 1274, 413, 532, 128, 373, 1142, 1882, 127, 1820, 1734, 1660, 1657, 1541, 1878, 1548, 1549, 1546, 1663, 1510, 1059, 1060, 1489, 1482, 1491, 1498, 1503, 983, 1496, 1507, 1001, 1493, 1735, 1508, 989, 1513, 1000, 1005, 1472, 1474, 1637, 1473, 1486, 1006, 1481, 993, 995, 1497, 1487, 1566, 1490, 1640, 1007, 1495, 992, 1470, 1471, 1488, 1739, 1740, 1003, 1501, 1509, 1639, 1478, 1564, 1505, 988, 1502, 987, 996, 1500, 999, 1004, 1476, 1477, 1479, 994, 1512, 985, 1484, 1504, 1641, 984, 1061, 1480, 1511, 1485, 990, 1494, 1475, 986, 1062, 1008, 1565, 1506, 1093, 1492, 997, 1058, 1738, 1736, 1737, 1638, 1002, 252, 1103, 973, 710, 838, 719, 1730, 1731, 22, 1906, 1255, 1861, 234, 282, 161, 1100, 1096, 1101, 1139, 1893, 1894, 1787, 1813, 1149, 268, 269, 716, 1628, 1904, 1705, 1706, 1098, 1099, 81, 465, 576, 1967, 1677, 1678, 1889, 1785, 1786, 1789, 1154, 1150, 1849, 1905, 966, 1974, 1630, 310, 731, 732, 280, 1151, 1254, 398, 1784, 184, 753, 1193, 272, 1148, 212, 553, 285, 258, 439, 1229, 1293, 233, 235, 717, 1783, 1230, 1292, 1294, 1257, 1258, 1598, 1751, 1750, 953, 1260, 1262, 213, 379, 922, 1212, 1213, 1840, 1699, 1209, 1210, 1596, 313, 934, 687, 565, 1847, 215, 1697, 394, 395, 1571, 1572, 946, 1835, 1746, 641, 1043, 1244, 1247, 892, 1871, 974, 1275, 1276, 921, 514, 887, 1267, 920, 1766, 399, 1407, 1343, 1344, 937, 1156, 791, 918, 926, 917, 1195, 709, 879, 1331, 160, 1308, 1522, 1728, 113, 15, 16, 18, 19, 915, 1778, 1808, 315, 1776, 1777, 1779, 1816, 1720, 1726, 1721, 914, 916, 957, 1804, 1406, 70, 624, 1333, 290, 1348, 1373, 1376, 1117, 980, 782, 1330, 1765, 289, 1792, 1802, 1817, 1818, 1727, 390, 1729, 1724, 956, 1644, 1645, 1389, 404, 502, 742, 1752, 1791, 1345, 157, 417, 83, 308, 1767, 642, 1398, 1145, 1606, 738, 1722, 254, 1397, 1781, 1346, 952, 150, 1379, 1380, 839, 1188, 1189, 1803, 919, 693, 1396, 85, 1146, 903, 209, 1347, 1304, 1306, 1307, 8, 1196, 1332, 1334, 1305, 629, 779, 1467, 1797, 739, 808, 1790, 166, 333, 350, 760, 1725, 585, 1335, 101, 1834, 1266, 14, 309, 1723, 1575, 1576, 194, 878, 780, 1268, 13, 963, 890, 893, 991, 244, 245, 246, 1719.
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Fiji, Finland, France, French Guiana, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Lao People's Democratic Republic, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Pakistan, Palestine, State of, Panama, Peru, Philippines, Poland, Portugal, R¿union, Romania, Russian Federation, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Viet Nam, Yemen.
986 units (156 US, 29 Canada, 801 ROW)