FDA Enforcement Class II Ongoing

BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM Pump Infusion Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 10013072; BD AlarisTM Bag Access Non-Vented Back Check Valve 2 SmartSiteTM Y-Sites, Catalog # 10013361; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 10013890; BD AlarisTM Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing SmartSiteTM Port (Burette) SmartSiteTM Y-Site, Catalog # 10015012; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE-Lined), Catalog # 10015862; BD AlarisTM Pump Infusion Blood Set 180 Micron Filter 2 Blunt Spikes, Catalog # 10062818; BD AlarisTM Pump Infusion Buerette Set SmartSiteTM Port (Burette) 3 SmartSiteTM Y-Sites, Catalog # 10821753; BD AlarisTM Pump Infusion Set 2 Back Check Valves 3 SmartSiteTM Y-Sites, Catalog # 11171447; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 11426964; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Back Check Valve Low Sorbing Tubing (PE Lined) 2 SmartSiteTM Y-sites, Catalog # 11532269; BD AlarisTM Pump Infusion Buerette Set 0.2Micron Filter Ball Valve SmartSiteTM Port (Burette) 2 SmartSiteTM Y-Sites, Catalog # 11613191; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 10015861A; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM ClosedMale LuerWith PrimingCAP, Catalog # 22000-B007T; BD AlarisTM Pump Infusion Set 1.2 Micron Filter, Catalog # 2202-0007; BD AlarisTM Pump Infusion Set, Catalog # 2204-0007; BD AlarisTM Pump Infusion Set Yellow-Striped Microbore Tubing, Catalog # 2206-0007; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 2260-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22600-0007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22602-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22603-B007T; BD AlarisTM Pump Infusion Set Back Check Valve SmartSiteTM Y-site, Catalog # 24001-0007; BD AlarisTM Pump Infusion Set Half Set SmartSiteTM Y-site, Catalog # 2403-0007; BD AlarisTMPumpInfusionSetSmartsiteTM BagAcessNon-Vented0.2MicronFilterBondedTexiumTMClosedLowSorbingTubing(PELined)MaleLuerwithprimingcapSmartSiteTMY-site, Catalog # 24301-0007T; BD AlarisTM Pump Infusion Buerette Set Smallbore Tubing SmartsiteTM Port (Burette) 3 SmartSiteTM Y-Site, Catalog # 2441-0007; BD AlarisTM Pump Infusion Set Back Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSiteTM Y-sites, Catalog # 2450-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 24600-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing TexiumTM Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 24601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-Site, Catalog # 2465-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined), Catalog # 2466-0007.

Recall: Z-0349-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0349-2024
Event ID
93266
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Becton Dickinson & Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 29, 2023
Initiation Date
October 11, 2023
Classification Date
November 17, 2023
Address
1 Becton Dr, N/A, Franklin Lakes, NJ, 07417-1815, United States

Description

BD Alaris Pump infusion sets (Product Name, Catalog #) BD AlarisTM Pump Infusion Set Vented syringe adapter Smallbore Tubing, Catalog # 10010483; BD AlarisTM Pump Infusion Set 1.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010453; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-site, Catalog # 10010454; BD AlarisTM Pump Infusion Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 10013072; BD AlarisTM Bag Access Non-Vented Back Check Valve 2 SmartSiteTM Y-Sites, Catalog # 10013361; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 10013890; BD AlarisTM Pump Infusion Buerette Set 0.2 Micron Filter Smallbore Tubing SmartSiteTM Port (Burette) SmartSiteTM Y-Site, Catalog # 10015012; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE-Lined), Catalog # 10015862; BD AlarisTM Pump Infusion Blood Set 180 Micron Filter 2 Blunt Spikes, Catalog # 10062818; BD AlarisTM Pump Infusion Buerette Set SmartSiteTM Port (Burette) 3 SmartSiteTM Y-Sites, Catalog # 10821753; BD AlarisTM Pump Infusion Set 2 Back Check Valves 3 SmartSiteTM Y-Sites, Catalog # 11171447; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 11426964; BD AlarisTM Pump Infusion Set 0.2 Micron Filter Back Check Valve Low Sorbing Tubing (PE Lined) 2 SmartSiteTM Y-sites, Catalog # 11532269; BD AlarisTM Pump Infusion Buerette Set 0.2Micron Filter Ball Valve SmartSiteTM Port (Burette) 2 SmartSiteTM Y-Sites, Catalog # 11613191; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 10015861A; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM ClosedMale LuerWith PrimingCAP, Catalog # 22000-B007T; BD AlarisTM Pump Infusion Set 1.2 Micron Filter, Catalog # 2202-0007; BD AlarisTM Pump Infusion Set, Catalog # 2204-0007; BD AlarisTM Pump Infusion Set Yellow-Striped Microbore Tubing, Catalog # 2206-0007; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined), Catalog # 2260-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22600-0007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22602-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined) Bonded TexiumTM with priming cap, Catalog # 22603-B007T; BD AlarisTM Pump Infusion Set Back Check Valve SmartSiteTM Y-site, Catalog # 24001-0007; BD AlarisTM Pump Infusion Set Half Set SmartSiteTM Y-site, Catalog # 2403-0007; BD AlarisTMPumpInfusionSetSmartsiteTM BagAcessNon-Vented0.2MicronFilterBondedTexiumTMClosedLowSorbingTubing(PELined)MaleLuerwithprimingcapSmartSiteTMY-site, Catalog # 24301-0007T; BD AlarisTM Pump Infusion Buerette Set Smallbore Tubing SmartsiteTM Port (Burette) 3 SmartSiteTM Y-Site, Catalog # 2441-0007; BD AlarisTM Pump Infusion Set Back Check Valve 2 Ganged 3-Way Stopcocks 3 SmartSiteTM Y-sites, Catalog # 2450-0500; BD AlarisTM Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 3 SmartSiteTM Y-Sites, Catalog # 24600-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing TexiumTM Back Check Valve 2 SmartSiteTM Y-sites, Catalog # 24601-B007T; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented 0.2 Micron Filter Low Sorbing Tubing (PE Lined) SmartSiteTM Y-Site, Catalog # 2465-0007; BD AlarisTM Pump Infusion Set SmartSiteTM Bag Access Non-Vented Low Sorbing Tubing (PE Lined), Catalog # 2466-0007.

Reason

Affected infusion sets labeled as free of DEHP actually do contain di(2-ethylhexyl) phthalate (DEHP).

Code Info

UDI-DI: 10885403232602, Catalog # 10010483; UDI-DI: 07613203021135, Catalog # 10010453; UDI-DI: 07613203015806, Catalog # 10010454; UDI-DI: 07613203021210, Catalog # 10013072; UDI-DI: 10885403221965, Catalog # 10013361; UDI-DI: 10885403232268, Catalog # 10013890; UDI-DI: 10885403233951, Catalog # 10015012; UDI-DI: 10885403221866, Catalog # 10015862; UDI-DI: 07613203019668, Catalog # 10062818; UDI-DI: 07613203019187, Catalog # 10821753; UDI-DI: 07613203021234, Catalog # 11171447; UDI-DI: 10885403232367, Catalog # 11426964; UDI-DI: 10885403232343, Catalog # 11532269; UDI-DI: 10885403276026, Catalog # 11613191; UDI-DI: 10885403232626, Catalog # 10015861A; UDI-DI: 10885403245022, Catalog # 22000-B007T; UDI-DI: 10885403274039, Catalog # 2202-0007; UDI-DI: 10885403199363, Catalog # 2204-0007; UDI-DI: 07613203019224, Catalog # 2206-0007; UDI-DI: 07613203012591, Catalog # 2260-0500; UDI-DI: 10885403239595, Catalog # 22600-0007T; UDI-DI: 10885403240966, Catalog # 22601-B007T; UDI-DI: 10885403240935, Catalog # 22602-B007T; UDI-DI: 10885403240942, Catalog # 22603-B007T; UDI-DI: 10885403238666, Catalog # 24001-0007; UDI-DI: 10885403232473, Catalog # 2403-0007; UDI-DI: 10885403223198, Catalog # 24301-0007T; UDI-DI: 10885403235047, Catalog # 2441-0007; UDI-DI: 10885403235115, Catalog # 2450-0500; ; UDI-DI: 10885403241970, Catalog # 24600-0007; UDI-DI: 10885403240959, Catalog # 24601-B007T; UDI-DI: 10885403221941, Catalog # 2465-0007; UDI-DI: 10885403221958, Catalog # 2466-0007. All lots, expiration date thru July 31, 2026.

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WI, WY, Puerto Rico, Guam; and the countries of Australia, Brazil, Canada, Dominican Republic, India, Kuwait, Mexico, New Zealand, Philippines, Poland, Saudi Arabia, Singapore, Taiwan, United Arab Emirates.

Quantity

N/A