FDA Enforcement Class III Terminated

A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis

Recall: Z-0349-2023 · Reported December 7, 2022

Enforcement

Recall Number
Z-0349-2023
Event ID
91054
Classification
Class III
Status
Terminated
Product Type
Devices
Firm
Primus Corporation dba Trinity Biotech
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 7, 2022
Initiation Date
October 5, 2021
Classification Date
December 1, 2022
Termination Date
August 9, 2023
Address
4231 E 75th Ter, Kansas City, MO, 64132-2059, United States

Description

A. MENARINI diagnostics Hb 9210 PREMIER, LEVEL1, LEVEL II, Hemoglobin A1c Controls for hemoglobin A1c Assay, REF 45052, part number 05-01-0926, A1c analysis

Reason

The Control Level I material value was showing results exceeding the upper limit of the acceptable range after a period of three (3) months. The unacceptable Control Level 1 value results in the Analyzer generating an error code which stops further testing and nullifies associated patient samples, resulting in a delay in patient results and a potential delay in therapeutic modifications.

Code Info

Lot 12161

Distribution

Turkey, India, Romania, Thailand, Vietnam, S. Korea, Peru, Pakistan Austria, UAE, Chile, Italy.

Quantity

340 kits