FDA Enforcement Class II Terminated

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Recall: Z-0344-2023 · Reported December 7, 2022

Enforcement

Recall Number
Z-0344-2023
Event ID
91178
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Mobius Imaging, LLC
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 7, 2022
Initiation Date
November 18, 2022
Classification Date
November 30, 2022
Termination Date
February 26, 2025
Address
2 Shaker Rd Ste F100, Shirley, MA, 01464-2535, United States

Description

MobiCT-32 AIRO Mobile CT Scanner, Cat. No. MobiCT-32, mobile computed tomography (CT) system, software version 2.1.1

Reason

Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.

Code Info

UDI-DI 00869346000200 Serial Numbers AIRO-0124, AIRO-0145, AIRO-0288

Distribution

Domestic distribution to consignees in Colorado, Tennessee, and Utah.

Quantity

3