FDA Enforcement Class II Terminated

Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following parent devices: Model 2030 (EPIC II), model 3005 S3 beds and stretcher. models 1105, 1105E, 1105X, 1105XE, 1115, 1115E, 1115X, 1115XE, 1125E, 1125X and 1125XE.

Recall: Z-0344-2016 · Reported December 9, 2015

Enforcement

Recall Number
Z-0344-2016
Event ID
72466
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Stryker Medical Division of Stryker Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 9, 2015
Initiation Date
October 13, 2015
Classification Date
November 30, 2015
Termination Date
September 6, 2016
Address
3800 E Centre Ave, Portage, MI, 49002-5826, United States

Description

Two-stage and three-stage I.V. poles are shipped as an optional accessory to a bed or stretcher. The I.V. poles are stationary, height adjustable stands intended to hold infusion liquids, infusion accessories, and/or other medical devices. They may be used on the following parent devices: Model 2030 (EPIC II), model 3005 S3 beds and stretcher. models 1105, 1105E, 1105X, 1105XE, 1115, 1115E, 1115X, 1115XE, 1125E, 1125X and 1125XE.

Reason

It was identified by a customer complaint that some I.V. poles were retracting/lowering from an extended position to the low height unexpectedly. An investigation was initiated which identified that the I.V. Latch Housing was nonconforming. potential hazards: " Suspended I.V. pole retracts unexpectedly " Suspended I.V. pole retracts unexpectedly with pump/bags attached

Code Info

Stryker 2-Stage and 3-Stage I.V. Poles distributed on beds, stretchers, and in service kits. This product field action is limited to I.V. Poles distributed between 01/07/2014 - 06/24/2014. The following I.V. Pole option numbers come affixed to beds and stretchers: Manufacturer Model Number and Product Description 2035-112-010; Head End Left, 2 Stage I.V. Pole 2035-113-011; Head End Right, 2 Stage I.V. Pole 0785-035-101; 26", 2 Stage IV Pole Assembly 0785-035-200; 26", 3 Stage IV Pole Assembly 0785-035-300; 30", 3 Stage IV Pole Assembly 0785-035-401; 30", 2 Stage IV Pole Assembly The following I.V. Pole option numbers ship as service kits. Manufacturer Service Kit Number and Product Description 1105-035-368; 30" IV POLE 2 STAGE HE PR 1105-035-367; 30" IV POLE 2 STAGE HE PL 1105-035-343; 30" IV POLE 2 STAGE FE PR 1105-035-341; 30" IV POLE, 2 STAGE HE, PR 1105-035-340; 30" IV POLE, 2 STAGE FE, PL 1105-035-338; 30" IV POLE, 2 STAGE FE, PL 1040-035-338; 2-STAGE IV POLEHEAD LT30" ZM 0785-035-340; 2-STAGE IV POLE,HEAD LT, 30" 0785-035-341; 2-STAGE IV POLEHEAD RT 30" 0785-035-343; 2-STAGE IV POLE,FOOT RT, 30" 1105-035-344; 30" IV POLE, 3 STAGE FE, PL 1105-035-342; 30" IV POLE, 3 STAGE HE, PL 1105-035-339; 30" IV POLE, 3 STAGE FE, PR 1105-035-369; 30" IV POLE, 3 STAGE HE, PL 1105-035-366; 30" IV POLE, 3 STAGE HE, PR 0785-035-339; 3-STAGE IV POLE,FOOT RT, 30" 0785-035-342; 3-STAGE IV POLE,HEAD LT, 30" 1105-035-644; 26" IV POLE, 3 STAGE FL 1105-035-642; 26" IV POLE, 3 STAGE HL 1105-035-639; 26" IV POLE 3 STAGE FR 1105-035-637; 26" IV POLE, 3 STAGE HR 0785-035-637; 3-STAGE IV POLE,HEAD RT, 26" 0785-035-639; 3-STAGE IV POLE,FOOT RT, 26" 0785-035-642; 3-STAGE IV POLE,HEAD LT, 26" 1105-035-643; 26" IV POLE, 2 STAGE FR 1105-035-641; 26" IV POLE, 2 STAGE HR 1105-035-640; 26" IV POLE, 2 STAGE FL 1105-035-638; 26" IV POLE, 2 STAGE HL 0785-035-638; 2-STAGE IV POLE,HEAD LT, 26" 0785-035-640; 2-STAGE IV POLE,FOOT LT, 26" 0785-035-641; 2-STAGE IV POLE,HEAD RT, 26" 0785-035-643; 2-STAGE IV POLE,FOOT RT, 26" 2035-112-000; IV POLE - 2 STAGE H/E 2035-113-000; DUAL H/E 2 STAGE IV POLE OPT

Distribution

Worldwide Distribution-US (nationwide) including DC and Guam, and the countries of New Zealand Canada Chile Hong Kong Colombia Argentina Brazil France Germany Italy Netherlands Spain Switzerland United Kingdom India Japan South Korea Mexico South Africa Poland Singapore.

Quantity

19,876