FDA Enforcement Class II Terminated

Grab n Go Opti series VIPR system Model # PRX-9627 - Product Usage: The Grab n Go Opti system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the device is for emergency use only. For all other medical applications, the device is Rx only. The device is MR Conditional and suitable for use outside the 1,500 gauss (0.15 T) line during MR imaging for MRI systems up to 3.0 T. The device is intended for limited duration use, such as would be necessary during patient transports.

Recall: Z-0340-2020 · Reported November 13, 2019

Enforcement

Recall Number
Z-0340-2020
Event ID
83857
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Western / Scott Fetzer Company
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Telephone
Report Date
November 13, 2019
Initiation Date
August 15, 2019
Classification Date
November 6, 2019
Termination Date
November 18, 2020
Address
875 Bassett Rd, N/A, Westlake, OH, 44145-1142, United States

Description

Grab n Go Opti series VIPR system Model # PRX-9627 - Product Usage: The Grab n Go Opti system is an integrated delivery system intended to provide supplemental oxygen, by Rx only, to patients. When administered by properly trained personnel for oxygen deficiency and resuscitation, the device is for emergency use only. For all other medical applications, the device is Rx only. The device is MR Conditional and suitable for use outside the 1,500 gauss (0.15 T) line during MR imaging for MRI systems up to 3.0 T. The device is intended for limited duration use, such as would be necessary during patient transports.

Reason

Limited access to flow settings as a result of the control knob having been rotated beyond its functional range

Code Info

All Lot manufactured from 08/31/14 to 08/07/2019

Distribution

Worldwide distribution  US Nationwide distributions in the states of OH, NY, PA, TX, MA, MI, KS, IL, MD, MN, WA, TN, UT, CA, NJ, LA, CT, OR, ND, RI, WI, PR, FL, ON, NV, NB, OK, IN, AZ, MS, GA, IA and countries of Costa Rica, Mexico.

Quantity

143,164 units