NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4
Recall: Z-0337-2016
·
Reported December 2, 2015
Enforcement
- Recall Number
- Z-0337-2016
- Event ID
- 72554
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- NxStage Medical, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 2, 2015
- Initiation Date
- October 29, 2015
- Classification Date
- November 25, 2015
- Termination Date
- November 19, 2019
- Address
- 350 Merrimack Street, N/A, Lawrence, MA, 01843-1748, United States
Description
NxStage System One S Cycler -High Permeability Hemodialysis System Model no. NX1000-4
Reason
Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed
Code Info
Software version 4.9 and 4.10
Distribution
Worldwide Distribution.
Quantity
7 units