FDA Enforcement Class II Terminated

Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electronic documentation of clinical data and is designed to accept, transfer, display, calculate, store and manage clinical data. System capabilities provide the user with the ability to acquire data from the medical devices and to document, annotate, display, store, print, export and retrieve patient clinical practice information. This product is intended for professional use only.

Recall: Z-0337-2014 · Reported November 27, 2013

Enforcement

Recall Number
Z-0337-2014
Event ID
64805
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
GE Healthcare It
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2013
Initiation Date
March 11, 2013
Classification Date
November 18, 2013
Termination Date
June 10, 2016
Address
540 W Northwest Hwy, Barrington, IL, 60010-3051, United States

Description

Centricity Perinatal and Centricity Intensive Care is intended to be used in clinical departments healthcare delivery systems such as a Labor and Delivery, Postpartum Maternal Care, newborn Nursery, critical care units including Neonatal Intensive Care Units (NICU), and may also be used in physicians offices and outpatient clinics. This product is primarily intended to serve the purpose of electronic documentation of clinical data and is designed to accept, transfer, display, calculate, store and manage clinical data. System capabilities provide the user with the ability to acquire data from the medical devices and to document, annotate, display, store, print, export and retrieve patient clinical practice information. This product is intended for professional use only.

Reason

1) Under rare network conditions, a Centricity Perinatal process may maintain an incorrect count of Patient slots in use which may lead to an inability to access the patient roster. The user is temporarily unable to review or edit form and chart information. This could result in a potential delay of care; however active fetal strips and permanent patient data storage are not affected. 2) When

Code Info

Centricity Perinatal and Centricity Intensive Care Versions: 6.60, 6.60.9, 6.70.0, 6.70.1, 6.70.5, 6.70.6, 6.80.0, 6.80.1, 6.90.0, 6.90.1, 6.91.1

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of: Australia, Canada, Denmark, Germany, Hong Kong, Korea, South Korea, Thailand and United Kingdom.

Quantity

1,132 units