FDA Enforcement Class II Terminated

NxStage System One S Cycler - Model no. NX1000-3-A (OUS). For home hemodialysis.

Recall: Z-0336-2016 · Reported December 2, 2015

Enforcement

Recall Number
Z-0336-2016
Event ID
72554
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
NxStage Medical, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 2, 2015
Initiation Date
October 29, 2015
Classification Date
November 25, 2015
Termination Date
November 19, 2019
Address
350 Merrimack Street, N/A, Lawrence, MA, 01843-1748, United States

Description

NxStage System One S Cycler - Model no. NX1000-3-A (OUS). For home hemodialysis.

Reason

Ultrafiltration Volume (UFV) may not decrease during treatment-software error. The UF pump may continue to run and remove fluid even after the target UF volume has been removed

Code Info

Software version 4.9 and 4.10

Distribution

Worldwide Distribution.

Quantity

133 units