Brilliance iCT Computed Tomography X-Ray System, Product Usage: The Brilliance iCT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Recall: Z-0335-2014 · Reported November 27, 2013

Enforcement

Recall Number
Z-0335-2014
Event ID
66601
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Philips Medical Systems (Cleveland) Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2013
Initiation Date
October 11, 2013
Classification Date
November 15, 2013
Termination Date
August 4, 2015
Address
595 Miner Rd, Cleveland, OH, 44143-2131, United States

Description

Brilliance iCT Computed Tomography X-Ray System, Product Usage: The Brilliance iCT is a Computed Tomography X-Ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason

If the couch, with the footrest extension attached, is positioned where the table travel is within the bore of the gantry, and the user begins to lower the couch, the footrest extension can contact the gantry cover or come in contact with the user due to the length of the accessory.

Code Info

Model #728306, Serial #'s: 100016, 100018, 100019, 100020, 100021, 100022, 100023, 100024 and 100025.

Distribution

Worldwide distribution: USA (nationwide) Distribution in the states of DC, GA, IL, IN, NY and OH and the countries: Germany, Israel and Netherlands.

Quantity

9 units