FDA Enforcement Class II Ongoing

Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.

Recall: Z-0334-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0334-2024
Event ID
93222
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Aizu Olympus Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Classification Date
November 17, 2023
Address
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, N/A, Japan

Description

Gastrointestinal videoscope-indicated for use within the upper digestive tract (including the esophagus, stomach, and duodenum). Model Numbers: GIF-1TH190, GIF-1TQ160, GIF-2T160, GIF-2TH180, GIF-H170, GIF-H180, GIF-H180J, GIF-H190, GIF-HQ190, GIF-LV1, GIF-N180, GIF-Q150, GIF-Q160Z, GIF-Q180, GIF-XP180N, GIF-XP190N, GIF-XTQ160. 510(k) Numbers: K112680, K954451, K051645, K100584, K111756, K011151, K123317.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Code Info

Model Numbers: GIF-1TH190 (04953170343360), GIF-1TQ160 (Not recorded), GIF-2T160 (04953170307492), GIF-2TH180 (04953170339745), GIF-H170 (04953170334115), GIF-H180 (04953170339769), GIF-H180J (04953170307539), GIF-H190 (04953170305290), GIF-HQ190 (04953170305276), GIF-LV1 (04953170312373), GIF-N180 (04953170307508), GIF-Q150 (04953170307485), GIF-Q160Z (04953170066795), GIF-Q180 (04953170307522), GIF-XP180N (04953170307515), GIF-XP190N (04953170305313), GIF-XTQ160 (04953170339837).

Distribution

Worldwide - US Nationwide distribution.

Quantity

3627 units