FDA Enforcement
Class II
Terminated
BD CD64 APC-R700; Analyte Specific Reagent; Model 657701; Lot No. 4344945, 5037611 and 5063945. Hematology - used as analyte specific reagent - contamination leads to possibility of unexpected staining pattern.
Recall: Z-0334-2016
·
Reported December 2, 2015
Enforcement
- Recall Number
- Z-0334-2016
- Event ID
- 72548
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Becton, Dickinson and Company, BD Biosciences
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- December 2, 2015
- Initiation Date
- October 30, 2015
- Classification Date
- November 24, 2015
- Termination Date
- May 27, 2016
- Address
- 2350 Qume Dr, N/A, San Jose, CA, 95131-1812, United States
Description
BD CD64 APC-R700; Analyte Specific Reagent; Model 657701; Lot No. 4344945, 5037611 and 5063945. Hematology - used as analyte specific reagent - contamination leads to possibility of unexpected staining pattern.
Reason
Three lots of CD64 (MD22) are contaminated with CD4 antibody.
Code Info
Model 657701; Lot No. 4344945, 5037611, both with expiry date of 31 DEC 2015; and 5063945 with expiry dte of 29 FEB 2016.
Distribution
Distributed in the states of FL, WA, PA, IL, GA, RI, TX, ND, and IA.
Quantity
23 vials