FDA Enforcement
Class II
Ongoing
CORMIX 150 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
Recall: Z-0332-2025
·
Reported November 20, 2024
Enforcement
- Recall Number
- Z-0332-2025
- Event ID
- 95499
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- The Metrix Company
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 20, 2024
- Initiation Date
- October 9, 2024
- Classification Date
- November 8, 2024
- Address
- 4400 Chavenelle Rd, N/A, Dubuque, IA, 52002-2655, United States
Description
CORMIX 150 mL EMPTY EVA CONTAINER With 2 Ports, REF MS64040; An empty single-use pouch made of plastic intended to contain a fluid for intravenous (IV) administration.
Reason
A limited number of IV bags have been found to leak during filling.
Code Info
UDI/DI 00812496011114, Lot Numbers: 64040-A8313, 64040-A8340, 64040-A8365, 64040-A8437
Distribution
Worldwide - US Nationwide distribution in the states of TX, IL, OH, PA, WA, MD, IL, UT, CA, MA, PA, MI, FL, PR, and the country of Canada.
Quantity
1920 cases of 50 units