FDA Enforcement Class II Terminated

ALTRUS¿ THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 36 cm Shaft, REF 60-9522-001, STERILE -- CONMED CORPORATION 525 FRENCH RD, UTICA, NY 13502 --- The ConMed Altrus¿ Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece. The Altrus¿ System utilizes a thermal energy platform to achieve the desired clinical effect. The ConMed Altrus¿ Thermal Tissue Fusion handpiece is a single use device which is provided sterile. The device uses scissors to parallel jaw closure mechanism with one flat jaw and one crowned jaw in which the vessels/tissues are grasped and through which pressure and heat are applied. It is a multi-functional device capable of vessel sealing, grasping and dissecting during open general surgical procedures. The two jaw sizes allow for applications in a variety of clinical environments for access in confined areas as well as large open areas.

Recall: Z-0332-2014 · Reported November 27, 2013

Enforcement

Recall Number
Z-0332-2014
Event ID
66547
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
ConMed Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 27, 2013
Initiation Date
October 23, 2013
Classification Date
November 15, 2013
Termination Date
March 8, 2017
Address
525 French Road, Utica, NY, 13502, United States

Description

ALTRUS¿ THERMAL TISSUE FUSION, Tissue Fusion Handpiece, 5 mm O.D., 36 cm Shaft, REF 60-9522-001, STERILE -- CONMED CORPORATION 525 FRENCH RD, UTICA, NY 13502 --- The ConMed Altrus¿ Thermal Tissue Fusion System is comprised of a dedicated energy source and disposable handpiece used to ligate (seal) and divide (cut) blood vessels and tissue bundles that fit into the jaws of the handpiece. The Altrus¿ System utilizes a thermal energy platform to achieve the desired clinical effect. The ConMed Altrus¿ Thermal Tissue Fusion handpiece is a single use device which is provided sterile. The device uses scissors to parallel jaw closure mechanism with one flat jaw and one crowned jaw in which the vessels/tissues are grasped and through which pressure and heat are applied. It is a multi-functional device capable of vessel sealing, grasping and dissecting during open general surgical procedures. The two jaw sizes allow for applications in a variety of clinical environments for access in confined areas as well as large open areas.

Reason

The affected lots were exposed to a higher than acceptable level of particulate matter during manufacture and subsequently distributed. ConMed evaluated the potential risk that this issue could present to the patient and concluded the risk of injury to be negligible; however exposure could cause a minor, local, self correcting infection.

Code Info

Lot 13CHB006

Distribution

Worldwide distribution: US (Nationwide) and countries of: Canada, Germany and Italy.

Quantity

147 handpieces (Domestic: 135 devices; Foreign: 12 devices)