FDA Enforcement Class III Ongoing

Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.

Recall: Z-0330-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0330-2024
Event ID
93222
Classification
Class III
Status
Ongoing
Product Type
Devices
Firm
Aizu Olympus Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Classification Date
November 17, 2023
Address
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, N/A, Japan

Description

Choleodochoscope. Model Numbers: CHF-BP30, CHF-CB30L, CHF-P60.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Code Info

Model Number (UDI-DI): CHF-BP30 (04953170068751), CHF-CB30L (04953170340154), CHF-P60 (Not recorded).

Distribution

Worldwide - US Nationwide distribution.

Quantity