FDA Enforcement Class II Terminated

Connection Screw for NCB Plate, Item No. 0202266002

Recall: Z-0330-2020 · Reported November 13, 2019

Enforcement

Recall Number
Z-0330-2020
Event ID
83962
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Zimmer Biomet, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 13, 2019
Initiation Date
October 10, 2019
Classification Date
November 7, 2019
Termination Date
May 19, 2020
Address
1800 W Center St, N/A, Warsaw, IN, 46580-2304, United States

Description

Connection Screw for NCB Plate, Item No. 0202266002

Reason

Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).

Code Info

12775518 12775518 12775527 12775527 12806273 12746627 12746627 12797742

Distribution

Distributed nationwide.

Quantity

5,459,583 total devices