FDA Enforcement
Class II
Ongoing
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Recall: Z-0329-2025
·
Reported November 20, 2024
Enforcement
- Recall Number
- Z-0329-2025
- Event ID
- 95490
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- WISSNER-BOSSERHOFF GMBH
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 20, 2024
- Initiation Date
- July 11, 2022
- Classification Date
- November 8, 2024
- Address
- Hauptstr. 4-6, N/A, Wickede (Ruhr), N/A, N/A, Germany
Description
SafeControl handset (a component of the Sentida bed which moves and positions the bed), Model Numbers: 01-011539, 01-011541, 01-01198
Reason
When the handset is disconnected from the power mains, and mains operation is restored, the handset is reset to the preset Resident mode. In this mode, the backrest, legrest, and height adjustment can be moved.
Code Info
UDI/DI 4050621100401, 19-0990 UDI/DI 4050621100012, 19-1062 UDI/DI 4050621100432, 19-1152 UDI/DI 4050621100449, 19-1119 UDI/DI 4050621100470, 19-1155 UDI/DI 4050621100487, 19-1156
Distribution
US Nationwide distribution in the states of NC.
Quantity
77 units