FDA Enforcement Class II Ongoing

Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.

Recall: Z-0328-2024 · Reported November 29, 2023

Enforcement

Recall Number
Z-0328-2024
Event ID
93222
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Aizu Olympus Co., Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 29, 2023
Initiation Date
September 25, 2023
Classification Date
November 17, 2023
Address
3 Chome 1-1, Niiderakita, Aizuwakamatsu, N/A, N/A, Japan

Description

Bronchoscope - Ultrasonic: endoscopic real-time ultrasonic imaging, ultrasound guided needle aspiration and other endoscopic procedures within the airways, trancheobronchial tree, esophagus and surrounding organs Model Numbers: BF-UC180F, BF-UC190F.

Reason

Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endoscopes has the potential for contamination that could lead to the risk of infection

Code Info

Model Number (UDI-DI): BF-UC180F (04953170356360), BF-UC190F (04953170399831).

Distribution

Worldwide - US Nationwide distribution.

Quantity