FDA Enforcement Class II Terminated

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Recall: Z-0328-2022 · Reported December 8, 2021

Enforcement

Recall Number
Z-0328-2022
Event ID
88999
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Randox Laboratories Ltd.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
December 8, 2021
Initiation Date
November 1, 2021
Classification Date
December 1, 2021
Termination Date
March 9, 2023
Address
55 Diamond Road, Crumlin (North), N/A, Ireland

Description

RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Reason

When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Code Info

Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Distribution

US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.

Quantity