FDA Enforcement
Class II
Terminated
RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200
Recall: Z-0328-2022
·
Reported December 8, 2021
Enforcement
- Recall Number
- Z-0328-2022
- Event ID
- 88999
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Randox Laboratories Ltd.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- December 8, 2021
- Initiation Date
- November 1, 2021
- Classification Date
- December 1, 2021
- Termination Date
- March 9, 2023
- Address
- 55 Diamond Road, Crumlin (North), N/A, Ireland
Description
RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200
Reason
When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range
Code Info
Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130
Distribution
US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.
Quantity
15 kits