FDA Enforcement Class I Terminated

Thermo SCIENTIFIC Sensititre Plate, GN4F, Gram Negative IVD AST

Recall: Z-0327-2023 · Reported December 14, 2022

Enforcement

Recall Number
Z-0327-2023
Event ID
91104
Classification
Class I
Status
Terminated
Product Type
Devices
Firm
Remel Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
December 14, 2022
Initiation Date
October 20, 2022
Classification Date
December 7, 2022
Termination Date
September 30, 2024
Address
12076 Santa Fe Trail Dr, N/A, Lenexa, KS, 66215-3519, United States

Description

Thermo SCIENTIFIC Sensititre Plate, GN4F, Gram Negative IVD AST

Reason

An internal technical investigation has confirmed that Thermo Scientific" Gram Negative IVD AST Sensititre Plate may lead to false susceptible results for Proteae Tribe (including Proteus spp., Providencia spp., Morganella Morganii) when tested with doripenem, ertapenem, meropenem, imipenem, cefepime, piperacillin/tazobactam and aztreonam.

Code Info

Lot Numbers: B2232B, B2197, B2012, B1355, B1181, B1173, B1125A

Distribution

US Nationwide Distribution

Quantity

1917 units