FDA Enforcement
Class II
Terminated
Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.
Recall: Z-0326-2017
·
Reported November 9, 2016
Enforcement
- Recall Number
- Z-0326-2017
- Event ID
- 75197
- Classification
- Class II
- Status
- Terminated
- Product Type
- Devices
- Firm
- Hill-Rom, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 9, 2016
- Initiation Date
- August 9, 2016
- Classification Date
- October 31, 2016
- Termination Date
- March 24, 2017
- Address
- 1069 State Road 46 E, Batesville, IN, 47006-7520, United States
Description
Product Number 3102071 Ceiling Bracket, Arch 71 White with use on ceiling installed rail systems: Multirall 200, Likorall 200, 242, 243, 250, and LikoGuard L and XL.
Reason
Ceiling Bracket 71 was shipped with a larger center hole. A potentially hazardous situation exists that if the Ceiling Bracket 71 with a larger center hole is used for a concrete ceiling Ultra System installation with the M8 screw and the safe working load is exceeded, the system could fall.
Code Info
manufacturing date between Dec 01, 2015- Apr 14, 2016
Distribution
WA NC OH PA MI Australia Sweden
Quantity
188