Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
Enforcement
- Recall Number
- Z-0325-2025
- Event ID
- 95559
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Baxter Healthcare Corporation
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 13, 2024
- Initiation Date
- October 11, 2024
- Classification Date
- November 7, 2024
- Address
- 1 Baxter Pkwy, N/A, Deerfield, IL, 60015-4625, United States
Description
Hillrom Welch Allyn XScribe Cardiac Stress Testing System; Programmable Diagnostic Computer
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Software Versions: Software Version V6.0.0 V6.3.2, Release Date 02-Jul-2015 Units: 1) UDI/DI 00732094259032, Serial Numbers: 115340269178, 117060553883, 117060553884, 117060553885, 119230003169; 2) UDI/DI 00732094259070, Serial Numbers: 117130559202, 117130559203, 117130559204; 3) UDI/DI 00732094259100, Serial Numbers: 116270497046, 117190567440; 4) UDI/DI 00732094259148, Serial Numbers: 117350583961, 119060000843, 122090000529; 5) UDI/DI 00732094259179, Serial Numbers: 116240458434, 116500543985, 120480000930; 6) UDI/DI 00732094259230, Serial Numbers: 118160628177; 7) UDI/DI 00732094259261, Serial Numbers: 117470596306; 8) UDI/DI 00732094259285, Serial Numbers: 118100618358, 116110348774, 116490541753, 117430591185, 117510601740; 9) UDI/DI 00732094259292, Serial Numbers: 116460537153, 118490000226; 10) UDI/DI 00732094259308, Serial Numbers: 116120365527, 119410001349, 119410001350, 119410001351, 119410001352, 119410001353, 119410001354, 119410001355, 119410001356, 119410001357, 119410001358, 119410001359, 119410001360, 119410001361, 119410001362, 119410001363, 119410001364, 119410001365, 119410001366, 119410001367, 119410001368, 119410001369, 119410001370; 11) UDI/DI 00732094259315, Serial Numbers: 117130559201; 12) UDI/DI 00732094259322, Serial Numbers: 117350583998, 117500601523; 13) UDI/DI 00732094259346, Serial Numbers: 116110347146, 116110347144; 14) UDI/DI 00732094259445, Serial Numbers: 116430533831, 116430533832; 15) UDI/DI 00732094259506, Serial Numbers: 116380527134, 117410589558, 118250001761, 119040000115, 119060000844, 119060000845, 123190000729, 123280000483, 123310000806, 123310000807, 123310000808, 123310000809, 123310000810, 123310000811, 123310000812, 123310000813, 123310000814, 123310000815, 123360000284, 123360000285, 123450001526, 123310000809, 123310000811, 123310000812, 123310000815; 16) UDI/DI 00732094259575, Serial Numbers: 118250001384, 118260000794, 118390001821, 118480000848, 118480000849, 119060000841, 119060000842, 119090000727, 119100000033, 119150000332, 119240000238, 119240000239, 119240000240, 119250000166, 120010000573, 120290000988, 121400001929, 121430000537, 122050000996, 122090000520, 122090000521, 122090000522, 122090000523, 122090000524, 122090000525, 122090000526, 122090000527, 122090000528, 122170001299, 123250001020, 123340000589, 123520000414, 124060000807, 118480000848, 124060000807, 124170000873, 124170000874, 124170000875, 124170000876, 124170000877, 124170000878, 124170000879, 124170000880, 124170000881, 124170000882, 124280000135, 124280000136, 124280000137, 124280000138, 124280000139, 124320001530; 17) UDI/DI 00732094259612, Serial Numbers: 116120365528; 18) UDI/DI 00732094259650, Serial Numbers: 116240458430, 117220570917, 118120622072, 118170632918, 118180634236, 118300000863, 118330000273, 118330000274, 118420000151, 118440001977, 118480000847, 118520000564, 119170000630, 119270000360, 119270000361, 119270000362, 119350003349, 119380001118, 119380001119, 119430000811, 119340002163, 119470001921, 120050001766, 120050001767, 120050001768, 120130001344, 120140002540, 120180001090, 120180001094, 120220000883, 120320001828, 120330000906, 120380000317, 120450001959, 120450001960, 120450001961, 120450001962, 120460002493, 120460002494, 121070000871, 121130001178, 121150000446, 121190000645, 121200001260, 121200001261, 121230001203, 121230001205, 121280001926, 121290000365, 121290000366, 121370002519, 121370002520, 121370002521, 121370002522, 121370002523, 121370002524, 121380000290, 121380000291, 121380000292, 121380000293, 121380000294, 121380000295, 121380000296, 121380000297, 121380000298, 121380000299, 121380000300, 122020001217, 122020001218, 122020001219, 122020001220, 122020001221, 122020001222, 122070000778, 122070000779, 122070000780, 122070000781, 122080000291, 122120002598, 122120002599, 122120002600, 122140000214, 122220001051, 122250001505, 122470000642, 122470000643, 122470000644, 122470000645, 122470000646, 122470000647, 122470000648, 122470000649, 122470000650, 122470000651, 123060000583, 123060000584, 123060000585, 123060000586, 123060000587, 123060000588, 123060000589, 123060000590, 123060000591, 123130000065, 123150000926, 123210000750, 123250001018, 123250001019, 123320000674, 123430000969, 124010000937, 124060000806, 120180001093, 121010002477, 123060000586, 123250001019, 123320000674, 123430000969, 124010000937, 124060000806, 124150000052, 124150000053, 124150000054, 124150000055, 124150000056, 124150000057, 124150000058, 124150000059, 124150000060, 124150000061, 124180000273, 124220000141, 124220000142; 19) UDI/DI 00732094259681, Serial Numbers: 118520000552, 118520000554, 119470001920, 121060000078, 121370000332, 122100000410, 122130000800, 122220000669, 122220000670; 20) UDI/DI 00732094259735, Serial Numbers: 118180635067, 118180635066, 118180635065; 21) UDI/DI 00732094259766, Serial Numbers: 118190636601, 118190636602, 118190636603, 118190636604, 118190636605, 118480000844, 118480000845; 22) UDI/DI 00732094259810, Serial Numbers: 118180635071, 118180635070, 118180635069, 118180635068, 119090000517, 119090000518, 122440001777, 124100001981, 124100001982; 23) UDI/DI 00732094259889, Serial Numbers: 117390588078; 24) UDI/DI 00732094259919, Serial Numbers: 116460538218, 116460538215, 116460538216, 116460538217; 25) UDI/DI 00732094259971, Serial Numbers: 118130624172, 118130624171, 118130624170, 118130624169; 26) UDI/DI 00732094260120, Serial Numbers: 116500544084, 119250000576, 120180001429; 27) UDI/DI 00732094260137, Serial Numbers: 117380586599, 117380586600; 28) UDI/DI 00732094260151, Serial Numbers: 117160563860, 117160563861, 117160563861; 29) UDI/DI 00732094260199, Serial Numbers: 117510602780, 117510602781; 30) UDI/DI 00732094260229, Serial Numbers: 116490541264, 116490541265, 117120557428, 117110556357, 117110556358, 117120557429, 119090000904, 119160000011, 121420001250, 121420001251, 121420001252, 122020000514, 122040000388, 122040000389, 122160000173, 122270001144, 122270001145, 122270001146, 122310000679, 122330000876, 122340000460, 122480000518, 123170000268; 31) UDI/DI 00732094260267, Serial Numbers: 116470539357, 116470539358, 116470539359, 116470539360, 116470539361, 116470539362, 116470539363, 116470539364, 116470539365, 116470539366, 116470539367, 116470539368, 118140625738, 119020000217, 119020000218, 119020000219, 119130001319, 119130001320, 119130001321, 119460001369, 121320001536, 123240001506, 123240001507, 124200000349; 32) UDI/DI 00732094260304, Serial Numbers: 117140562699, 117140562700; 33) UDI/DI 00732094260342, Serial Numbers: 116510545162; 34) UDI/DI 00732094260380, Serial Numbers: 116070308454, 116480540034, 118300000033, 118300000034, 118300000035, 118300000036, 118300000037, 118300000038, 118300000039, 118300000041, 118300000042, 118300000043, 118300000044, 118300000045, 118300000046, 118300000047, 118300000048, 118300000049, 118300000050, 118300000051, 118300000052, 118300000053, 118300000054, 118300000055, 118340000935, 118340000936, 118340000937, 118340000938, 118340000939, 118340000940, 118340000941, 118340000942, 118340000943, 118460001733, 118460001734, 119240000094, 120190001967, 120190001968, 120190001969, 120220001542, 120220001543, 120220001544, 120220001545, 120220001546, 122070001255, 118300000053, 118300000055, 120220001542; 35) UDI/DI 00732094260434, Serial Numbers: 118300000860, 118460001730, 118460001732, 118480000841, 119130000112, 119130000113, 119130000114, 119200001280, 119200001614, 119380001489, 121150000434, 121150000435, 121360000218, 121450000733, 121520000986, 121520000987, 121520000988, 121520000989, 119380001489; 36) UDI/DI 00732094260519, Serial Numbers: 113030149726, 113030149727, 118120623345, 114160206692, 115170254512, 115170254513, 117270576959, 118230000780, 118240001065, 118240001637, 118300000859, 118330001333, 118440000169, 118480000837, 119060000839, 119090000902, 119170000481, 119170000482, 119170000483, 119170000484, 119170000485, 119200000583, 119200001823, 119230002791, 119250000575, 119380001481, 119380001482, 119490000072, 119490000073, 120220000881, 120220000882, 120290000977, 120290000978, 120520000902, 121230000437, 121250001830, 122050000686, 122050000687, 122070000773, 122070000774, 122070000775, 122070000776, 122070000777, 120440001019, 120440001013, 122350000281, 122440001772, 122440001773, 122500001549, 122500001550, 123100001477, 123160001934, 123160001935, 123230001128, 123100001477, 123160001934, 123160001935, 123200001637, 123230001128, 124160002135, 124160002136, 124170000872, 124180000269, 124220001303, 124220001304; 37) UDI/DI 00732094260533, Serial Numbers: 117390588115, 117390588116, 117390588117, 117390588118, 111130075526; 38) UDI/DI 00732094260557, Serial Numbers: 114150206492, 114450235208, 117390588112, 117390588113, 117390588114, 118480000835, 119350003345, 121280001920, 121280001921, 122070001253, 122070001254, 122130000799, 122260000202, 122260000203, 122260000204, 122260000205, 122260000206, 122260000207, 122260000208, 123070001809, 123090001127, 123110000416, 123200001005, 123450001334, 123450001335, 114150206492, 123090001127, 123110000416, 123450001334, 123450001335, 124110001068, 124110001069, 124160000521; 39) UDI/DI 00732094260601, Serial Numbers: 113480185300; 40) UDI/DI 00732094260670, Serial Numbers: 118100617600, 113100154916, 118130624595, 118330000155, 118330000156, 118480000834, 119180000558, 119250001418, 119260000959, 120080000601, 120460002490, 121130001169, 121130001170, 122070001249, 122070001250, 122070001251, 122070001252, 122290000263, 122290000264, 122290000265, 122340000461, 122340000462; 41) UDI/DI 00732094260731, Serial Numbers: 116120368670; 42) UDI/DI 00732094260816, Serial Numbers: 113240166924, 117080554664, 117080554665, 117220571840, 118280000973, 118480000833; 43) UDI/DI 00732094260847, Serial Numbers: 117520603150; 44) UDI/DI 00732094260885, Serial Numbers: 116340523128, 116350524044. 117370586263; 45) UDI/DI 00732094260892, Serial Numbers: 117380586444; 46) UDI/DI 00732094260960, Serial Numbers: 117470596307, 118050609226; 47) UDI/DI 00732094260991, Serial Numbers: 116500543942, 116500543943, 116500543944, 116500543945, 116500543946; 48) UDI/DI 00732094261004, Serial Numbers: 116390529064; 49) UDI/DI 00732094261028, Serial Numbers: 116460538216, 116460538217, 116110348790, 116070311448, 117370586263, 116070311445, 116070311446, 116070311447, 119380001488; 50) UDI/DI 00732094261042, Serial Numbers: K16C001871, K16C001464, 118140625485, 118140625484, 118130623667, 118120623366, 118120623365, 118120623364, 118120623363, 118120623362, 118120623360, 118120623359, 118120623356, 118120623355, 118120623358, 118120623361, 118120623357, 118110619537, 118110619536, 118140625486, 118140625488, 118140625489, 118140625490, 118140625491, 118140625492, 118140625493, 119050000696, 119350003340, 119470002302, 119490000509, 21119350003340, 21119490000509, 119520000628, 120010000572, 120050001447, 21120010000572, 21120050001447, 121010000496, 118120623358; 51) UDI/DI 00732094261196, Serial Numbers: 118270000496, 118270000497, 118270000498, 118310000442, 118310000443, 118310000444, 118310000445, 118310000446, 118310000447, 118460002316; 52) UDI/DI 00732094294255, Serial Numbers: 118280001663, 118280001664, 118450000678, 118450000679, 118280001663; 53) UDI/DI 00732094294538, Serial Numbers: 120140002538, 120140002539, 120320001826, 120320001827, 121060001018, 122220000671; 54) UDI/DI 00732094294545, Serial Numbers: 122170001364; 55) UDI/DI 00732094295429, Serial Numbers: 118360000421, 118370002214; 56) UDI/DI 00732094296303, Serial Numbers: 119420001730, 119420001731, 119420001732, 119420001733, 119420001734, 119420001735, 119420001736, 121510000646, 121510000647, 121500000303, 123180000483, 123180000484; 57) UDI/DI 00732094299908, Serial Numbers: 11810623357; 58) UDI/DI 00732094299915, Serial Numbers: 118120623355, 118120623356, 118120623365, 118120623359, 118120623363; 59) UDI/DI 00732094301472, Serial Numbers: 117500600621, 120440000589, 120440000590, 120440000591, 120440000592, 120440000593, 122440001762, 21120440000592, 12244001762, 122470000639; 60) UDI/DI 00732094302974, Serial Numbers: 119020000643, 121320001539, 121320001540, 121320001541, 12341, 12342, 12343, 121420001245, 121420001246, 121420001247, 121420001248; 61) UDI/DI 00732094316926, Serial Numbers: 119140001089, 119460000150, 119460000151; 62) UDI/DI 00732094320176, Serial Numbers: 122360001053, ....; 63) UDI/DI 00732094320183, Serial Numbers: 120110001003, 121050001860; 64) UDI/DI 00732094332780, Serial Numbers: 117470596305, 120070001961, 120070001962, 120070001963, 120070001964, 120070001965, 120070001966, 120070001967, 120070001968, 120070001969, 120070001970, 120070001971, 120070001972, 120070001973, 120070001974, 120070001975, 120070001976, 120070001977, 120070001978, 120070001979, 120070001980, 120070001965, 120070001970, 120070001978; 65) UDI/DI 00732094337518, Serial Numbers: 115350269606, 120270000572, 121470000326; 66) UDI/DI 00732094338805, Serial Numbers: 114340225721; 67) UDI/DI 00732094339444, Serial Numbers: 120400001336; 68) UDI/DI 00812345027556, Serial Numbers: 116130382889, 116250489751, 116490541979, 117260576365, 118260000798, 118260000799, 117270577497, 117270577498, 118400001533, 118460001735, 119100000089, 119100000090, 120450001963, 120530000477, 121140001840, 121250000689, 121380000301, 121400001503, 121500001615, 121500001616, 122020001911, 122020001912, 122100000411, 122130000801, 122130000802, 122130000803, 122370001278, 122390000503, 122390000504, 122390000505, 122480002479, 122510000876, 123080000426, 123080000427, 123170000296, 123520001479, 123520001480, 124030000954, 123520001474, 121380000301, 124030000954, 124170000105, 124190001474; 69) UDI/DI 00817655021997, Serial Numbers: 117370585639
Worldwide distribution.
766 units