FDA Enforcement Class II Ongoing

Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system

Recall: Z-0323-2026 · Reported November 5, 2025

Enforcement

Recall Number
Z-0323-2026
Event ID
97715
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Surgical Theater Inc
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
E-Mail
Report Date
November 5, 2025
Initiation Date
September 16, 2025
Classification Date
October 24, 2025
Address
23645 Mercantile Rd Ste M, N/A, Beachwood, OH, 44122-5936, United States

Description

Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system

Reason

A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.

Code Info

UDI/DI 00863157000329

Distribution

Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.

Quantity

19 units