FDA Enforcement Class II Ongoing

HURRICANE RX DILATION BALLOON 6MM 2CM-for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN:M00545910

Recall: Z-0322-2022 · Reported December 8, 2021

Enforcement

Recall Number
Z-0322-2022
Event ID
88991
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Report Date
December 8, 2021
Initiation Date
October 21, 2021
Classification Date
December 1, 2021
Address
100 Boston Scientific Way, Marlborough, MA, 01752-1234, United States

Description

HURRICANE RX DILATION BALLOON 6MM 2CM-for endoscopic dilatation of strictures of the biliary tree and the Sphincter of Oddi. UPN:M00545910

Reason

Balloon pinhole results in losss pressure or fails to gain or maintain pressure, causing prolonged procedure

Code Info

GTIN: 8714729283829 Batches: 25623385, 25678822, 25739151, 25771772, 25975774, 26010392, 26051928, 26087733, 26123186, 26175469, 26406521, 26429810, 26456904, 26579839, 26581406, 26601326, 26660957, 26793786, 26896864, 27026985, 27096335, 27226570, 27255234, 27400203. Exp. 05/23/2022 to 05/28/2023

Distribution

US Nationwide distribution.