FDA Enforcement Class II Terminated

Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures

Recall: Z-0322-2017 · Reported November 9, 2016

Enforcement

Recall Number
Z-0322-2017
Event ID
75124
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Medtronic Navigation, Inc.
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 9, 2016
Initiation Date
September 1, 2016
Classification Date
October 28, 2016
Termination Date
March 27, 2017
Address
826 Coal Creek Cir, Louisville, CO, 80027-9710, United States

Description

Visualase Visualization Stylets PN 020-2301. The individual components within the Visualase Body Accessory Kit are packaged in Tyvek pouches. The packaged devices are then placed into a cardboard box which is labeled with the Body Accessory Kit label. Labeled sterile. The visualization stylet is constructed of titanium and used during Visualase laser ablation procedures by inserting it into intracorporeal catheter lumens for improved catheter stiffness and MRI visualization during procedures

Reason

Medtronic Navigation, Inc. announces a voluntary field action for the Visualase Visualization Stylets and Visualase Body Accessory Kit because when the Visualization Stylet is placed in the magnetic field of a magnetic resonance imaging system it pulls and/or bends toward the field.

Code Info

Lot number 151481

Distribution

Worldwide Distribution - US including CA, GA, FL, and TX. Internationally to Australia.

Quantity

100 total units