FDA Enforcement
Class II
Ongoing
Surgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system
Recall: Z-0320-2026
·
Reported November 5, 2025
Enforcement
- Recall Number
- Z-0320-2026
- Event ID
- 97715
- Classification
- Class II
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Surgical Theater Inc
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Report Date
- November 5, 2025
- Initiation Date
- September 16, 2025
- Classification Date
- October 24, 2025
- Address
- 23645 Mercantile Rd Ste M, N/A, Beachwood, OH, 44122-5936, United States
Description
Surgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system
Reason
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
Code Info
UDI/DI 00863157000305
Distribution
Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Quantity
17 units