FDA Enforcement Class II Ongoing

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

Recall: Z-0319-2024 · Reported November 22, 2023

Enforcement

Recall Number
Z-0319-2024
Event ID
93362
Classification
Class II
Status
Ongoing
Product Type
Devices
Firm
Philips North America
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
Letter
Report Date
November 22, 2023
Initiation Date
October 9, 2023
Classification Date
November 16, 2023
Address
222 Jacobs St, N/A, Cambridge, MA, 02141-2296, United States

Description

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

Reason

This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).

Code Info

UDI-DI: 00884838103627; Lot Code: 1005

Distribution

US Nationwide distribution in the state of MN.

Quantity

1 unit