FDA Enforcement Class II Terminated

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

Recall: Z-0319-2021 · Reported November 11, 2020

Enforcement

Recall Number
Z-0319-2021
Event ID
86534
Classification
Class II
Status
Terminated
Product Type
Devices
Firm
Boston Scientific Corporation
Voluntary / Mandated
Voluntary: Firm initiated
Initial Notification
N/A
Report Date
November 11, 2020
Initiation Date
September 22, 2020
Classification Date
October 31, 2020
Termination Date
September 4, 2024
Address
4100 Hamline Ave N Bldg 3, N/A, Saint Paul, MN, 55112-5700, United States

Description

Model 3300 LATITUDE Programming System with installed Model 3892 ALTRUA, INSIGNIA I, NEXUS I Software Support Application

Reason

There is potential when a user changes an EGM trace channel with a Manual Pace Threshold test in progress that the newly modified trace channel may change to a flat line.

Code Info

Model 3892 v1.04 is installed on Model 3300 LATITUDE Programming Systems with the following serial numbers: 009494 012689 010356 008206 008124 007983 004813 010751

Distribution

Distributed in US - AR, FL, IL, and TX

Quantity

8 devices