Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
Enforcement
- Recall Number
- Z-0318-2026
- Event ID
- 97607
- Classification
- Class I
- Status
- Ongoing
- Product Type
- Devices
- Firm
- Dexcom, Inc.
- Voluntary / Mandated
- Voluntary: Firm initiated
- Initial Notification
- Letter
- Report Date
- November 5, 2025
- Initiation Date
- August 28, 2025
- Classification Date
- October 30, 2025
- Address
- 6340 Sequence Dr, N/A, San Diego, CA, 92121-4356, United States
Description
Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System
A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).
Lot Code: UDI: SW11678 - 00386270000811
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.
333